Emergency Care Equipment<br />

GCE Gas Control Equipment AB<br />

Skeppsbron 3<br />

211 20 Malmö<br />

Sweden<br />

Tel: +46 40-38 83 45<br />

Fax: +46 40-38 83 33<br />

www.gceab.se<br />

GCE Norden AB<br />

FlygfäLtsvägen 1<br />

212 39 Malmö<br />

Sweden<br />

Tel: +46 40-38 83 00<br />

Fax: +46 40-38 83 50<br />

www.gcenorden.se<br />

GCE Ltd<br />

Yew Tree Way<br />

Stone Cross, Golbourne<br />

Cheshire<br />

Warrington W<br />

U.K.<br />

Tel: +44 1695-714474<br />

Fax: +44 1695-50925<br />

www.gceuk.com<br />

GCE S.A.<br />

6, rue de Gérigny<br />

BP No 110<br />

58404 La Charité-sur-Loire<br />

Cedex<br />

France<br />

Tel: +33 386-69 46 00<br />

Fax: +33 386-70 09 15<br />

www.gcesa.fr<br />

GCE MUJELLI S.p.A<br />

Via F.lli Cervi, 11<br />

37038 S.Martino B.A.<br />

Verona<br />

Italy<br />

Tel: +39 045-87 80 525<br />

Fax: +39 045-87 80 750<br />

www.mujelli.it<br />

GCE Hungária Kft<br />

H-2314 Halásztelek<br />

ll. Rákóczi Ferenc út 90/B<br />

Hungary<br />

Tel: +36 24-521 200<br />

Fax: +36 24-521 201<br />

www.gce.hu<br />

GCE Sp z o.o.<br />

Jagiellonska 74<br />

03-301 Warsawa<br />

Poland<br />

Tel: +48 22-675 27 67<br />

Fax: +48 22-675 29 80<br />

www.gce.pl<br />

GCE Autogen s.r.o.<br />

Zizkova 381<br />

583 81 Chotebor<br />

The Czech Republic<br />

Tel: +420 569-66 11 11<br />

Fax: +420 569-66 16 021<br />

www.gce.cz<br />

GCE-RHÖNA GmbH<br />

Postfach 1328<br />

36039 Fulda<br />

Germany<br />

Tel: +49 661-83 930<br />

Fax: +49 661-83 93 31<br />

www.gce-rhoena.de<br />

GCE Ibérica<br />

Calle Juan de Olias 21, 1<br />

28020 Madrid<br />

Spain<br />

Tel: +34 915 71 14 70<br />

Fax: +34 915 71 27 56<br />

www.gcesa.fr<br />

GCE Romania SRL<br />

104E Timisoara Bvd.<br />

6 Bucharest<br />

Rumania<br />

Tel: +40 1 444 09 76<br />

Fax: +40 1 444 09 77<br />

GCE Inc.<br />

4457 Bethany Road, Building H<br />

Manson, OH 450 40<br />

USA<br />

Tel +1 513 315 8801<br />

Fax +1 513 339 1083<br />

<strong>Instruktion</strong><br />

<strong>Demandventil</strong> <strong>för</strong> <strong>oxygen</strong><br />

<strong>Instruction</strong> <strong>of</strong> <strong>use</strong><br />

<strong>Oxygen</strong> Demand Valve<br />

GCE mediline is a trading name <strong>use</strong>d by operation companies within<br />

the GCE Group, the parent company <strong>of</strong> which is GCE Holding AB,<br />

Sweden<br />

<strong>Instruktion</strong> / <strong>Instruction</strong> <strong>of</strong> <strong>use</strong> 700 326 006<br />

050418ao


S<br />

Innehåll<br />

sid<br />

1 Inledning<br />

1.1 Allmänt 2<br />

1.2 Tekniska data 2<br />

1.3 Termer och defi nitioner 2<br />

1.4 Säkerhets<strong>för</strong>eskrifter 2<br />

1.5 Garanti 2<br />

2 Tillvägagångssätt<br />

2.1 Spontanandning 3<br />

2.2 Bruksanvisning 3<br />

3 Service<br />

3.1 Rutinunderhåll 3<br />

3.2 Underhåll 3<br />

3.3 Rengöring av demandventilen 4<br />

3.4 Tillvägagångssätt vid rengöringen 4<br />

4 Reservdelar och tillbehör till Mediline ventilator<br />

med demandventil 4<br />

5 Återvinning 4<br />

1. Inledning<br />

1.1 Beskrivning<br />

• Mediline demandventil <strong>för</strong> <strong>oxygen</strong> är ett säkert och effektivt hjälpmedel <strong>för</strong> utbildad<br />

personal <strong>för</strong> att till<strong>för</strong>a <strong>oxygen</strong> vid inandning.<br />

• Mediline demandventil <strong>för</strong> <strong>oxygen</strong> är lätt, bärbar och mycket hållbar. Den har utformats<br />

i enlighet med de krav som ställs vid användning i prehospital vård.<br />

1.2 Tekniska data<br />

Manuellt fl öde: 0-160 liter/min ± 10%<br />

Tryckkänslighet <strong>för</strong> ventilöppning: 100 liter/min @ < -5 cm H 2<br />

O ±10%<br />

Minimalt ingångstryck: 3 bar<br />

Maximalt utgångstryck: 5 bar<br />

Tryckavlastning :<br />

45 cm H 2<br />

O<br />

Övertryckslarm:<br />

konstant ljudsignal<br />

Drifts-/<strong>för</strong>varingstemperatur: -37 °C till +50 °C<br />

Ingångsanslutning: 9/16”<br />

Patientanslutning: 15/22 mm<br />

Vikt: 227 g<br />

1.3 Termer och definitioner<br />

• <strong>Demandventil</strong>: En ventil som levererar gas till patienten med en fl ödes- hastighet<br />

motsvarande den som indikeras av patientens inandning.<br />

• Utandningsfas: Tiden från slutet på inandningsfl ödet till slutet på utandnings- fl ödet.<br />

• Utandningstid: Den tid utandningen varar.<br />

• <strong>Oxygen</strong>koncentration: <strong>Oxygen</strong>innehållet i de gaser som inandats uttryckt i procent.<br />

• Minutvolym (Vm): Den totala gasvolym som levereras till patienten under en minut.<br />

• Tryckavlastningsventil: Ventil som begränsar lungans maximala utvidgnings- tryck<br />

genom att tömma ut överfl ödsgas till atmosfär.<br />

• Andningsvolym (Vt): Den gasvolym som levereras till patienten under varje inandningsfas.<br />

1.4 Säkerhets<strong>för</strong>eskrifter<br />

Mediline demandventil <strong>för</strong> <strong>oxygen</strong> är avsedd att användas av lämpligt utbildad personal.<br />

Följande <strong>för</strong>siktighetsåtgärder bör alltid iakttas:<br />

1. Rök inte när apparaten är i bruk. Vistas inte heller nära öppen låga vare sig då apparaten<br />

används eller vid fl askbyte.<br />

2. Stäng alltid gasfl askan när apparaten inte används.<br />

8 Ensure that the cylinder is turned OFF after each <strong>use</strong> and remaining gas released<br />

from the system by slightly loosening the regulator/demand valve connection until<br />

gas fl ow ceases. Always retighten the connection after bleeding the surplus pressure.<br />

3 Servicing<br />

3.1 Routine maintenance<br />

NOTE:<br />

The Mediline <strong>oxygen</strong> demand valve is designed to provide years <strong>of</strong> reliable service in<br />

all emergency situations. Failure to follow the maintenance and inspection routines<br />

properly could result in incorrect operation <strong>of</strong> the demand valve.<br />

To ensure proper operation <strong>of</strong> the demand valve regular inspection and checking <strong>of</strong><br />

the demand valve and accessories for correct function should be undertaken by a<br />

responsible member <strong>of</strong> staff on a regular basis.<br />

This check is to ensure that all <strong>of</strong> the accessories and demand valve components are<br />

present, the gas cylinder is full and that the demand valve is in working order.<br />

3.2 Maintenance<br />

Regulator working pressure should be checked at least every six months and more<br />

frequently in high <strong>use</strong> applications. Units with test pressures outside <strong>of</strong> the ranges<br />

listed in the product specifi cations should not be <strong>use</strong>d. The Mediline products are not<br />

designed for fi eld disassembly or service outside that indicated in this manual. Any<br />

malfunctioning units should be returned to the manufacturer. Unauthorized repairs will<br />

nullify the product warranty.<br />

3.3 Cleaning the demand valve<br />

Routine cleaning <strong>of</strong> the Mediline <strong>oxygen</strong> demand valve should be undertaken to<br />

maintain the equipment in a clean condition. Face masks, patient valve and diaphragm<br />

can all be disinfected using a legally marketed commercially available<br />

disinfectant, steam sterilization or ethylene oxide gas. All other components should be<br />

cleaned with a mild soap solution.<br />

NOTE: Do not attempt to clean and sterilize any components that are designated as<br />

disposable, as immersion <strong>of</strong> these items into a sterilizing solution can ca<strong>use</strong> degeneration<br />

<strong>of</strong> the materials.<br />

3.4 Cleaning procedure<br />

The <strong>oxygen</strong> demand valve must be thoroughly cleaned after each patient <strong>use</strong>.<br />

1 Ensure that the cylinder is turned <strong>of</strong>f (completely clockwise) and any<br />

remaining gas bled out <strong>of</strong> the regulator.<br />

2 Remove the face mask from the <strong>oxygen</strong> demand valve and then disconnect<br />

the demand valve from the regulator.<br />

3 Unscrew patient valve from the demand valve body being careful to ensure<br />

that the diaphragm is retained.<br />

4 Wash all components thoroughly in a mild soap solution and disinfect as<br />

required.<br />

5 The handpiece can be wiped over with a s<strong>of</strong>t cloth and mild soap solution.<br />

6 Dry all components thoroughly.<br />

7 Re-assemble and check performance.<br />

4 The Mediline demand valve replacement parts<br />

325 111 852 Patient valve diaphragm<br />

325 111 884 Hose nut, nipple with O-ring<br />

325 111 886 O-ring for hose nipple<br />

325 111 855 Universal face mask<br />

5 Recycling<br />

Metal parts shall be recycled by scrapping, plastic parts and rubber details can be<br />

2 7


phase.<br />

1.4 Safety precautions<br />

The Mediline <strong>oxygen</strong> demand valve is intended for <strong>use</strong> by suitably trained and qualifi<br />

ed personnel. The following precautions should always be observed:<br />

1 When the unit is in <strong>use</strong>, do not smoke or <strong>use</strong> near open fl ame either during <strong>use</strong> or<br />

when changing the cylinder.<br />

2 When not in <strong>use</strong>, always turn <strong>of</strong>f the gas cylinder.<br />

3 Never allow oil or grease to come into contact with any part <strong>of</strong> the cylinder, regulator<br />

or demand valve.<br />

4 Do not disassemble any part <strong>of</strong> the demand valve except where described in this<br />

manual as any unauthorized disassembly will invalidate the warranty.<br />

5 After <strong>use</strong>, always ensure that all components are cleaned in accordance with the<br />

instructions provided in this manual.<br />

6 Always <strong>use</strong> the check list to ensure that all components are reassembled correctly<br />

and ready for <strong>use</strong>.<br />

7 After <strong>use</strong>, always ensure that a full gas cylinder is attached before returning the unit<br />

to its normal storage position.<br />

1.5 Warranty<br />

The Mediline <strong>oxygen</strong> demand valve equipment is manufactured from the fi nest<br />

quality materials. Each individual part is subject to strict quality control tests to ensure<br />

exceptionally high standards. The manufacturer warrants to the purchaser <strong>of</strong> the<br />

Mediline <strong>oxygen</strong> demand valve that its component parts are free from defects in<br />

material and workmanship for a period <strong>of</strong> two years from the date <strong>of</strong> purchase. The<br />

manufacturer will replace and/or repair all parts <strong>of</strong> the <strong>oxygen</strong> demand valve at its<br />

option for two years from the date <strong>of</strong> purchase at no cost to the purchaser, upon the<br />

notifi cation <strong>of</strong> the defects, in writing by the purchaser. All shipping costs shall be borne<br />

by the purchaser. The manufacturer shall be liable under this warranty only if the Mediline<br />

<strong>oxygen</strong> demand valve and its parts have been <strong>use</strong>d and serviced in the normal<br />

manner described in the instruction manual. There are no other expressed or implied<br />

warranties. This warranty gives no specifi c legal rights.<br />

2 Operating procedure<br />

2.1 Demand breathing<br />

The Mediline <strong>oxygen</strong> demand valve is a device powered by any 3-5 bar regulated<br />

<strong>oxygen</strong> supply. Its function is to provide gas to the patient at a safe delivery pressure<br />

and adequate fl ow rate upon demand.<br />

An inspiratory effort by the patient will open the demand valve and <strong>oxygen</strong> will fl ow to<br />

the patient at a rate in line with their inspiratory effort.<br />

2.2 Operating instructions<br />

1 Ensure that the <strong>oxygen</strong> demand valve/regulator connection is tight.<br />

2 Turn the cylinder handwheel counter clockwise to the ON position (at least one full<br />

turn).<br />

3 Ensure that the regulator contents gauge reads greater than 50 bar. If not, replace<br />

with a full cylinder.<br />

4 Attach the universal face mask to the demand valve outlet.<br />

5 Instruct the patient in its operation.<br />

6 To trigger the demand valve and obtain an adequate fl ow <strong>of</strong> gas, it is necessary<br />

to make a good mask to face seal. Simply place the mask on the patient’s face,<br />

covering both the nose and mouth. Instruct the patient to hold the mask on his/her<br />

face with a light pressure and breathe normally. The sound <strong>of</strong> gas fl owing should be<br />

heard with each inspiration. If not, adjust the patient’s hold on the mask for a better<br />

seal and ask the patient to breathe more deeply.<br />

7 Continue administration as per local management protocol.<br />

3. Låt aldrig olja eller fett komma i kontakt med någon del av fl askan, regulatorn eller<br />

demandventilen.<br />

4. Demontera inte någon del av demandventilen utom då så beskrivs i denna bruksanvisning,<br />

eftersom varje icke auktoriserad demontering kommer att göra garantin<br />

ogiltig.<br />

5. Kontrollera alltid efter användning att alla komponenter har rengjorts i enlighet med<br />

anvisningarna i denna bruksanvisning.<br />

1.5 Garanti<br />

Mediline demandventil <strong>för</strong> <strong>oxygen</strong> tillverkas av de fi naste kvalitetsmaterial. Varje<br />

enskild del är <strong>för</strong>emål <strong>för</strong> strikt kvalitetskontroll <strong>för</strong> att säkerställa en utomordentligt hög<br />

standard. Köparen av Mediline demandventil <strong>för</strong> <strong>oxygen</strong> erhåller tillverkarens garanti<br />

att utrustningens komponenter vad avser material och tillverkning kommer att vara<br />

felfria under en period om två år från inköpsdatum. Tillverkaren ersätter och/eller<br />

reparerat; efter eget val, alla delar av demandventilen <strong>för</strong> <strong>oxygen</strong> under två år från<br />

inköpsdatum utan kostnad <strong>för</strong> köparen, efter skriftligt påpekande från köparen om<br />

felen. Alla fraktkostnader skall betalas av köparen. Tillverkaren ansvarar <strong>för</strong> denna<br />

garanti endast om Mediline demandventil <strong>för</strong> <strong>oxygen</strong> och dess delar har använts och<br />

servats på normalt sätt som beskrivs i denna instruktion. Inga övriga uttryckta eller under<strong>för</strong>stådda<br />

garantier fi nns. Denna garanti med<strong>för</strong> inga specifi ka lagliga rättigheter.<br />

2 Tillvägagångssätt<br />

2.1 Spontanandning<br />

• Mediline demandventil <strong>för</strong> <strong>oxygen</strong> drivs med 3-5 bar reglerad <strong>oxygen</strong>till<strong>för</strong>sel. Dess<br />

uppgift är att leverera gas till patienten med säkert leveranstryck och lämplig fl ö-<br />

deshastighet i <strong>för</strong>hållande till spontanandningen.<br />

• Vid patientens inandnings <strong>för</strong>sök öppnas demandventilen och <strong>oxygen</strong> fl ödar till<br />

patienten med en hastighet som motsvarar inandningen.<br />

2.2 Bruksanvisning<br />

1. Kontrollera att anslutningen <strong>oxygen</strong>demandventilenlregulatorn är tät.<br />

2. Kontrollera att regulatorns innehållsmätare står på mer än SO bar. Om inte, byt till<br />

en full fl aska.<br />

3. Anslut universalmasken till demandventilens utgång.<br />

4. Instruera patienten om hur apparaten fungerar.<br />

5. För att demandventilen skall kunna sättas igång och ett lämpligt gas fl öde erhållas<br />

måste masken täta ordentligt till ansiktet. Masken placeras helt enkelt på patientens<br />

ansikte så att den täcker både näsa roch mun. Be patienten hålla masken till<br />

ansiktet med ett lätt tryck samt andas normalt. Ljudet av gas som fl ödar bör höras<br />

vid varje inandning. Om så inte är fallet, justera patientens grepp så att masken<br />

tätar bättre.<br />

6. Efter varje användning skall överbliven gas släppas ut ur systemet genom att man<br />

frigör snabbkopplingshandelen från snabbkopplingen på regulatorn.<br />

3 Service<br />

3.1 Rutinunderhåll<br />

OBS!<br />

• Mediline demandventil <strong>för</strong> <strong>oxygen</strong> har utformats <strong>för</strong> att ge många års till<strong>för</strong>litlig<br />

service i alla akutlägen. Underlåtenhet att följa underhålls- och kontrollrutinerna<br />

ordentligt kan resultera i att demandventilen inte fungerar på rätt sätt.<br />

• För att demandventilen säkert ska fungera ordentligt bör kontroller och funktionstester<br />

göras regelbundet av ansvarig personal.<br />

• Dessa kontroller skall göras <strong>för</strong> att <strong>för</strong>säkra sig om att demandventilens alla tillbehör<br />

och komponenter fi nns på plats, att gasfl askan är full samt att demandventilen är<br />

funktionsduglig.<br />

6 3


3.2 Underhåll<br />

Regulatorns arbetstryck skall kontrolleras åtminstone var sjätte månad och ännu<br />

<strong>of</strong>tare vid frekvent användning. Apparater vars provtryckvärden hamnar utan<strong>för</strong> de i<br />

produktspecifi kationerna angivna värdena skall inte användas. Mediline-produkterna<br />

är inte avsedda att demonteras på fältet eller att servas på annat sätt än som anges i<br />

denna bruksanvisning. Alla apparater som inte fungerar skall returneras till tillverkaren.<br />

Icke auktoriserade reparationer gör produktgarantin ogiltig.<br />

3.3 Rengöring av demandventilen<br />

Mediline demandventil <strong>för</strong> <strong>oxygen</strong> skall rengöras rutinmässigt.<br />

• Ansiktsmasker, patientventil och membran kan alla desinfi ceras genom att man använder<br />

ett i handeln legalt marknads<strong>för</strong>t desinfektionsmedel, sterilisering med ånga<br />

eller etylenoxidgas.<br />

• Alla övriga komponenter skall rengöras med en mild tvållösning.<br />

OBS!<br />

Försök inte rengöra eller sterilisera några komponenter som betecknas som engångsartiklar,<br />

eftersom nersänkning av sådana komponenter i en steriliseringslösning kan<br />

med<strong>för</strong>a att materialet blir <strong>för</strong>stört.<br />

3.4 Tillvägagångssätt vid rengöringen<br />

Mediline demandventil <strong>för</strong> <strong>oxygen</strong> måste rengöras grundligt efter varje patientanvändning.<br />

1. Kontrollera att gasfl askan är avstängd.<br />

2. Ta bort ansiktsmasken från demandventilen <strong>för</strong> <strong>oxygen</strong> och frikoppla därefter demandventilen<br />

från regulatorn.<br />

3. Skruva loss patientventilen från demandventilens hus och var därvid noga med att<br />

kontrollera att membranet är helt.<br />

4. Tvätta alla komponenter grundligt i mild tvållösning och desinfi cera efter behov.<br />

5. Själva ventilh<strong>use</strong>t kan torkas av med en mjuk duk och mild tvållösning.<br />

6. Torka alla komponenter ordentligt.<br />

7. Montera ihop igen och kontrollera funktionen.<br />

4 Reservdelar till Mediline demandventil<br />

325 111 852 Membran till patientventil<br />

325 111 886 O-ring <strong>för</strong> slangnippel<br />

325 111 855 Universalmask<br />

325 112 059 Patientventil<br />

5 Återvinning<br />

Metalldelar skall återvinnas genom skrotning, plastdelar och gummidetaljer kan <strong>för</strong>brännas.<br />

Table <strong>of</strong> contents<br />

page<br />

1 Introduction<br />

1.1 General description 5<br />

1.2 Technical data 5<br />

1.3 Terms and defi nitions 5<br />

1.4 Safety precautions 6<br />

1.5 Warranty 6<br />

2 Operating procedure<br />

2.1 Demand breathing 6<br />

2.2 Operating instructions 6<br />

3 Servicing<br />

3.1 Routine maintenance 7<br />

3.2 Maintenance<br />

3.3 Cleaning the demand valve 7<br />

3.4 Cleaning procedure 7<br />

4 The Mediline demand valve replacement parts 7<br />

5 Recycling 7<br />

1. Introduction<br />

1.1 General description<br />

The Mediline <strong>oxygen</strong> demand valve provides trained individuals with a safe and effective<br />

means <strong>of</strong> delivering <strong>oxygen</strong> by inhalation.<br />

The Mediline <strong>oxygen</strong> demand valves are lightweight, portable, and extremely durable.<br />

Designed for the demands <strong>of</strong> the pre-hospital and in-hospital environment, they can be<br />

utilized anywhere a regulated <strong>oxygen</strong> supply is available.<br />

1.2 Technical data<br />

Demand flow rate: 0-160 litres/min ±10%<br />

Demand fl ow triggering pressure: 100 litres/min @ < -5 cmH 2<br />

O ±10%<br />

Minimum input pressure:<br />

3 bar<br />

Maximum output pressure:<br />

5 bar<br />

Pressure relief:<br />

45 cmH 2<br />

O<br />

Overpressure alarm:<br />

audible constant<br />

Operating/storage temp:<br />

-37 o C to - +50 o C<br />

Inlet connection: 9/16”<br />

Patient connection:<br />

15/22 mm<br />

Weight:<br />

227 g<br />

1.3 Terms and definitions<br />

Airway resistance: Pressure drop across the airway per unit fl ow.<br />

Demand valve: A valve that delivers gas to the patient at a fl owrate equivalent to<br />

that demanded by the patient’s inspiratory effort.<br />

Expiratory phase: The time period from the end <strong>of</strong> the inspiratory fl ow to the end <strong>of</strong> the<br />

expiratory fl ow.<br />

Expiratory time (TE): Duration <strong>of</strong> the expiratory phase.<br />

<strong>Oxygen</strong> concentration: The <strong>oxygen</strong> content <strong>of</strong> the inspired gases expressed as a<br />

percentage.<br />

Lung compliance: Volume added per unit pressure increase when gas is added to a<br />

human or artifi cial lung.<br />

Minute volume (Vm): The total volume <strong>of</strong> gas delivered to the patient in one minute.<br />

Pressure relief valve: Valve which limits the maximum lung infl ation pressure by venting<br />

excess gas to atmosphere.<br />

Tidal volume (Vt): Volume <strong>of</strong> gas delivered to the patient during each inspiratory<br />

GB<br />

4 5