Lamp replacement<br />
Starter replacement<br />
8.3.2 Starter replacement (P/N 100-1370)<br />
• Turn off the power.<br />
• Disconnect the power cord from the Universal Hood II.<br />
• Remove the four screws located on the left/right sides of the drawer.<br />
• Remove the cover with the UV filter by sliding it forward.<br />
• Place it down on its backside on a non-abrasive surface.<br />
Caution: Do not put the UV filter directly on the bench.<br />
• To remove the starter, rotate it counter-clockwise and pull out.<br />
• To replace the starter insert it into the holder and rotate clockwise.<br />
• Reassemble the cover and retighten the screws on both sides.<br />