Lamp replacement<br />

Starter replacement<br />

8.3.2 Starter replacement (P/N 100-1370)<br />

• Turn off the power.<br />

• Disconnect the power cord from the Universal Hood II.<br />

• Remove the four screws located on the left/right sides of the drawer.<br />

• Remove the cover with the UV filter by sliding it forward.<br />

• Place it down on its backside on a non-abrasive surface.<br />

Caution: Do not put the UV filter directly on the bench.<br />

• To remove the starter, rotate it counter-clockwise and pull out.<br />

• To replace the starter insert it into the holder and rotate clockwise.<br />

• Reassemble the cover and retighten the screws on both sides.<br />