This unit is protected by 2 fuses 5X20 mm 2A Slow Blow. The fuses are located in the rear left<br />

side in a fuse holder. See above picture.<br />

1. Unplug the main power cable from the power entry module outlet.<br />

2. Using a screwdriver lever on the fuse holder to extract it.<br />

3. Remove the blown fuses and replace them with two new ones (P/N 100-1951).<br />

4. Slide the fuse holder into the power entry module until it snaps in place<br />

8.3 UV Transilluminator<br />

8.3.1 Lamp replacement<br />

UV transilluminator without cover<br />

Attention: The UV filter surface should always be kept clean from the chemical agents<br />

used as gel dyes. Please use protective gloves when touching the UV transilluminator<br />

cover.<br />

The lamps must be replaced after every 500 hours of use.<br />

• Turn off the power.<br />

• Disconnect the power cord from the Universal Hood II.<br />

• Remove the four screws located on the left/right sides of the drawer.<br />

• Remove the cover with the UV filter by sliding it forward.<br />

• Place it down on its backside on a non-abrasive surface.<br />

Caution: Do not put the UV filter directly on the bench.<br />

• Wear gloves when touching the lamps. Remove the lamp by rotating it until it becomes<br />

loose and the pins come to a vertical position;<br />

• Remove the lamp. Install the new lamp by rotating so that the pins are horizontal and the<br />

lamp is tight.<br />

• Reassemble the cover and retighten the screws on both sides.<br />

Page 33 of 42