Section 8 Maintenance and Part Replacement<br />
This section covers the replacement of parts in the Universal Hood II.<br />
8.1 Epi-illumination lamp replacement<br />
The lamps are located behind the two panels on each side (left and right) of the internal side of<br />
the hood.<br />
• Turn system power off.<br />
• Remove the power cord.<br />
• Open the enclosure door.<br />
• Locate the Epi lamp housing.<br />
• Locate the socket head screw that hold each lamp housing to the internal side of the hood.<br />
• Remove this screw.<br />
• The cover screen with a plastic piece will come loose. Pull on the cover to remove it from<br />
the box and the lamp will become visible.<br />
• To remove the lamp, hold it from the receptacle and then pull it from the plastic Holder.<br />
• Insert the new lamp (5 or 11 W) into the lamp holder and then push it into the receptacle.<br />
• Reassemble the cover.<br />
8.2 Fuse replacement<br />
Epi-White Light 5 W assembly (left side)<br />
Always unplug the instrument before changing or checking the fuses.<br />
Power Cable plug-in<br />
Fuse holder<br />
ON/OFF Main Switch<br />
Power Entry Module<br />