Section 8 Maintenance and Part Replacement<br />

This section covers the replacement of parts in the Universal Hood II.<br />

8.1 Epi-illumination lamp replacement<br />

The lamps are located behind the two panels on each side (left and right) of the internal side of<br />

the hood.<br />

• Turn system power off.<br />

• Remove the power cord.<br />

• Open the enclosure door.<br />

• Locate the Epi lamp housing.<br />

• Locate the socket head screw that hold each lamp housing to the internal side of the hood.<br />

• Remove this screw.<br />

• The cover screen with a plastic piece will come loose. Pull on the cover to remove it from<br />

the box and the lamp will become visible.<br />

• To remove the lamp, hold it from the receptacle and then pull it from the plastic Holder.<br />

• Insert the new lamp (5 or 11 W) into the lamp holder and then push it into the receptacle.<br />

• Reassemble the cover.<br />

8.2 Fuse replacement<br />

Epi-White Light 5 W assembly (left side)<br />

Always unplug the instrument before changing or checking the fuses.<br />

Power Cable plug-in<br />

Fuse holder<br />

ON/OFF Main Switch<br />

Power Entry Module<br />

Page 32 of 42