Note: When using the White Light, you must turn ON the UV Transilluminator after<br />
removing the sample.<br />
12. The system will automatically acquire a reference image and a flat-fielded corrected image<br />
of the sample will be generated for you to save and analyze.<br />
13. For additional details on Flat Fielding please refer to the Quantity One® Software Manual.<br />
You may also click on the Quantity One® HELP button for a detailed description.<br />
5.2 Cutting gels<br />
1. Open the drawer until it is fully extended and the UV Trans LED stops blinking.<br />
2. Place the clear acrylic UV protection screen in the horizontal slot in the inside of the<br />
drawer facing the transilluminator.<br />
!! CAUTION!! !! VORSICHT !!<br />
WARNING: EXPOSURE TO UV RADIATION IS HAZARDOUS TO HEALTH.<br />
PLEASE WEAR PROPER UV PROTECTIVE CLOTHING AND FACE AND EYE<br />
SHIELDS WHEN CUTTING GELS. THE UV SHIELD PROVIDED WITH THE<br />
SYSTEM IS NOT ADEQUATE PROTECTION AGAINST UV GENERATED BY<br />
THIS SYSTEM.<br />
3. Press the UV Trans button to turn on the UV illumination.<br />
4. Cut the gel.<br />
5. Press the UV Trans button and the UV transilluminator will turn off. Remove the acrylic<br />
UV protection shield before closing the drawer.<br />