STEP 2: Initialization of the Camera and lens assembly Procedure (for <strong>Gel</strong> Doc <strong>XR</strong>/<br />
ChemiDoc <strong>XR</strong>S systems):<br />
1. Press the Epi-Illumination button.<br />
2. Open the door and determine if lights are on.<br />
3. Place the focusing target on the UV transilluminator.<br />
4. Double click on Quantity One® software icon on the desktop. From “File Menu” select<br />
“<strong>Gel</strong> Doc <strong>XR</strong>” or “ChemiDoc <strong>XR</strong>S”.<br />
NOTE: Dark Reference comes up on first use for both <strong>Gel</strong> Doc <strong>XR</strong> and ChemiDoc<br />
<strong>XR</strong>S systems. Please refer to next step for details about “Reference Dark” image<br />
acquisition.<br />
5. Select Live/Focus.<br />
6. An image of the target will appear on the computer screen. Using the lens control buttons<br />
on the membrane touch pad or the control panel that appears in the software, find the best<br />
Iris, Focus, and Zoom conditions.<br />
Note: Working with a widely open Iris and changing the Zoom may require you to<br />
adjust the Focus.<br />
STEP 3: Dark Reference Acquisition:<br />
Note: Please note that the steps described below must be performed in either of the<br />
following scenarios:<br />
a. When installing the system with a new computer or Camera<br />
b. When software is first installed or reinstalled for any reason<br />