2. ENVIRONMENTAL ChEMISTRy & TEChNOLOGy 2.1. Lectures
2. ENVIRONMENTAL ChEMISTRy & TEChNOLOGy 2.1. Lectures
2. ENVIRONMENTAL ChEMISTRy & TEChNOLOGy 2.1. Lectures
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Chem. Listy, 102, s265–s1311 (2008) Environmental Chemistry & Technology<br />
between acidified and non-acidified samples. Similarly, changes<br />
in retention times were negligible. The remaining issue<br />
with the use of MES was the increase in baseline noise and<br />
subsequent increase in detection limits. This problem was<br />
overcome by pre-cleaning the buffer using a column packed<br />
with Eichrom Diphonix ® resin. The resulting baseline noise<br />
reduced approximately three times and the corresponding<br />
background fluorescence was almost seven times less.<br />
It was thus determined that a pre-cleaned buffer of<br />
0.25M MES adjusted to a pH of 6.05 with naOH, was the<br />
optimum choice for the determination of aluminium in acidified<br />
seawater samples.<br />
Temperature<br />
The response of the reaction between aluminium and<br />
lumogallion has been investigated in both batch techniques<br />
and flow systems. In the batch method, an optimal temperature<br />
of 80 °C is generally accepted 13,15 , whereas FIA methods<br />
tend to use 50 °C. The latter is based on investigations carried<br />
out by Resing and Measures which concluded that most of<br />
the temperature-based reaction rate gain had been achieved<br />
by this temperature 12 . Independent investigation into the effect<br />
of temperature on the rate of reaction was undertaken by<br />
us due to the fact a different buffer was used. It was found<br />
that the highest response, in terms of peak area, was obtained<br />
at temperatures between 65 and 75 °C (Fig. 1.). Based on this<br />
response, 70 °C was chosen as the temperature at which to<br />
operate the post column reactor for all subsequent anlyses.<br />
Fig. 1. Dependence of fluorescence response on temperature<br />
Lumogallion and Reaction Coil<br />
The extent of chemical reaction needs not be complete<br />
for an analytical technique to be valid. However, it is desirable<br />
to obtain as high a reaction yield as possible in order<br />
to ensure the technique has good precision. For the reaction<br />
between aluminium and lumogallion, the concentration of<br />
post-column reagent may be changed, along with temperature<br />
and reaction time, in order to control the extent of reaction.<br />
Three concentrations of lumogallion (0.03, 0.04 and<br />
s321<br />
0.05 mM) were tested in order to exhaust possible improvements<br />
to the system via this approach. The concentrations<br />
chosen were based on those used in flow systems. It was<br />
found that at concentrations higher than 0.03 mM, no significant<br />
improvements were achived. Additionally, the effect of<br />
increasing the length of the post column reaction coil from<br />
2m to 4m was also studied. The result, however, was a slight<br />
reduction in fluorescence. A MES buffer containing 0.03 mM<br />
lumogallion together with a 2m reaction coil were thus used<br />
in all subsequent analyses.<br />
Surfactant<br />
Howard and co-workers reported an increase in the fluorescence<br />
intensity of the aluminium-lumogallion complex<br />
of as much as 5-fold through the addition of a non-ionic<br />
surfactant 13 . Further investigation has been carried out by<br />
Resing and Measures 12 , which showed that Brij-35 enhanced<br />
fluorescence to a greater extent than other surfactants, such as<br />
Triton X-100 and cetylammonium bromide (CTAB). In order<br />
to ensure the lowest limit of detection was achieved for this<br />
system, an investigation into the effect of surfactants was also<br />
carried out. The results differed substantially from those discussed<br />
earlier. It was found that although the addition of Brij-<br />
35 enhanced fluorescence marginally, a simultaneous increase<br />
in baseline noise negated any improvement achieved.<br />
Interestingly, when CTAB was tested, the aluminium peak<br />
disappeared altogether. This was considered to be an effect<br />
of the surfactant adhering to the tubing walls and effectively<br />
stripping the aluminium from the reagent stream. The system<br />
required flushing with methanol in order to resume normal<br />
operation. Consequently, further investigation into the<br />
possible use of surfactants was abandoned, with the decision<br />
to explore other approaches to lowering the detection limit<br />
being deemed more favourable.<br />
S a m p l e V o l u m e<br />
A more attractive approach for achieving a low LOD<br />
was increasing the sample loop volume. All previous experiments<br />
had been carried out using a volume of 20 µl. The<br />
response of the system to higher volumes was investigated<br />
and the results are depicted in Fig. <strong>2.</strong> It can be seen that for<br />
volumes between 20 and 500 µl, the system follows a linear<br />
response, as expected. It was also noteworthy that no reduction<br />
in column efficiency was experienced at higher volumes.<br />
The highest efficiency was achieved for a 100 µl sample<br />
loop, which was unexpected considering that band broadening<br />
is generally associated with increased sample size and<br />
is often responsible for an observed reduction in performance<br />
of the chromatographic column as injection volume is increased.<br />
Another unexpected result of increasing the sample<br />
volume was an increase in retention time. Generally, a decrease<br />
in retention time would be expected due to competition<br />
from other analytes for chelation sites, especially in such a<br />
complex matrix as seawater. This was shown not to be the<br />
case for IDAS and may be explained in terms of the forma-