5.2 Semiconductor Spectrometers<br />

- 20 -<br />

W. Michaelis, F, Horsch, D. Lange, 0. Meyer, TI. Tamm<br />

5.2.1 Detector Technology<br />

Coaxial Li-drifted germanium detectors have been fabricated with<br />

active volumes up to 33 cm £1J r . The energy resolution (3.3 keV<br />

for Bcv = 662 keV) is comparable with the energy resolution of planar<br />

detectors. The time distribution has been investigated for various<br />

detector shapes (circular and trapezoidal cross section, one open<br />

end and double-open end detectors).<br />

High-resolution particle detectors are now produced by ion-implantation<br />

technique £"2.J* The improved performance of these counters is due<br />

to the fact that well-channeled ions at relative low energy and low<br />

ion-concentration minimize bulk damage effects. High-resistivity<br />

n-type, p-type and intrinsic germanium and silicon have been bombar­<br />

ded with ions of more than 30 elements in the energy range between<br />

+ +<br />

5 and 10 keV ¿ 3.y. The properties of large-area n and p contacts<br />

produced in this way have been studied in detail. Very thin entrance<br />

windows are obtained by ion-implantation.<br />

The gamma-ray response characteristics of lithium-drifted germanium<br />

detectors have been determined experimentally for several detector<br />

geometries J^^J*<br />

5*2.2 Electronic Instrumentation<br />

For high-resolution studies preamplifiers with cooled paralleled<br />

field-effect transistors have been applied with great success £~5_7.<br />

A pulse-shape discrimation circuit has been developed which is<br />

sensitive to a slow time-constant component in the charge carrier<br />

collection of lithium-drifted germanium detectors /J~6_7. Pre­<br />

amplifier output pulses containing such a component are rejected.<br />

By this method the spectrum shapes can be improved considerably.