New product focus: Ultra View Temperature (& Humidity) Chamber<br />
3<br />
Table 6 shows the main specifications of the new chambers.<br />
Table 6 Main specifications<br />
Model PWL-2KP PWL-3KP PWL-4KP PWU-2KP PWU-3KP PWU-4KP<br />
Power supply<br />
Temperature (& humidity)<br />
control system<br />
- 28 -<br />
200V AC 3Ф 3W 50/60Hz,<br />
220V AC 3Ф 3W 60Hz,<br />
380V AC 3Ф 4W 50 Hz<br />
Balanced temperature and humidity<br />
Balanced temperature control (BTC)<br />
control (BTHC)<br />
Working temperature range 0 to +40°C<br />
Temperature range -40 to +120°C<br />
Humidity range 20 to 98%rh ----------<br />
Temperature<br />
-40 to<br />
+100°C<br />
±0.3°C<br />
fluctuation +100.1 to<br />
+120°C<br />
±0.5°C<br />
Humidity fluctuation ±2.5%rh ----------<br />
Performance<br />
*1<br />
Temperature<br />
-40 to<br />
+100°C<br />
±0.5°C ±1.0°C ±0.5°C ±1.0°C<br />
uniformity +100.1 to<br />
+120°C<br />
±0.75°C ±1.5°C ±0.75°C ±1.5°C<br />
Humidity uniformity ±3.0%rh ±5.0%rh ----------<br />
Temperature<br />
heat-up time<br />
Within 50 min from -40 to +120°C<br />
Temperature<br />
pull-down time<br />
Within 60 min from +20 to -40°C<br />
Effective viewing dimensions<br />
(W×H mm)<br />
Dimension<br />
*2<br />
Specifications<br />
Test area dimensions<br />
(W×H×Dmm)<br />
Outer dimensions<br />
(W×H×D mm)<br />
470x720 570x820 970x970 470x720 570x820 970x970<br />
500×<br />
705×<br />
600<br />
910×<br />
1590×<br />
1039<br />
600×<br />
850×<br />
800<br />
1010×<br />
1690×<br />
1239<br />
1000×<br />
1000×<br />
800<br />
1410×<br />
1970×<br />
1239<br />
500×<br />
705×<br />
600<br />
910×<br />
1590×<br />
1039<br />
600×<br />
850×<br />
800<br />
1010×<br />
1690×<br />
1239<br />
1000×<br />
1000×<br />
800<br />
1410×<br />
1970×<br />
1239<br />
Loading capacity (L) 225 408 800 225 408 800<br />
Weight (kg) 310 370 560 300 360 550<br />
* 1:・Performance values are given for +23°C ambient temperature and no specimens<br />
inside chamber.<br />
・Temperature/Humidity fluctuation and uniformity are based on JTM (Standard of<br />
Testing Machinery Association of Japan) K 01-1998.<br />
・Loading capacity is given for an effective area up to 1/6 the distance from walls in<br />
all direction.<br />
* 2: Excluding projections.<br />
Espec Technology Report No20