download
download
download
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
New product focus: Ultra View Temperature (& Humidity) Chamber<br />
2-3 Markedly improved performance<br />
Using conventional heat-reflecting film or incorporating heating wires in the glass panes<br />
required design limits on the performance of the chambers, but by utilizing glass<br />
reinforced with layers of deposit metal film, performance can be markedly upgraded. This<br />
design permits a wide temperature range, from –40°C to +120°C.<br />
Table 4 Comparison of lowest and highest achievable temperatures<br />
Lowest achievable emperature Highest achievable temperature<br />
New chambers -40°C *1 +120°C *2<br />
Established chambers -20°C to -35°C +80°C to +100°C<br />
*1 This performance table conforms to JTM K01-1998.<br />
*2 With the optional “viewing window reach-in ports”, the limit is +100°C.<br />
Expanding this temperature range makes it possible to apply the following types of test<br />
conditions as well.<br />
<br />
(1)LCD device environmental tests<br />
Storage (at high temperature) test: +30 to +100°C<br />
Storage (at low temperature) test*: -50 to 0°C<br />
Damp heat, steady state test: +40°C, 93%rh to +85°C, 93%rh; 85°C, 85%rh<br />
Composite temperature/humidity cyclic test: +25°C, 93%rh to +65°C, 93%rh;<br />
-10°C<br />
Standards: IEC61747-5 (JIS C60068-2-38), EIAJ ED-2531A<br />
(2)Automotive parts environmental tests<br />
High Temperature, Low Temperature Test: +120°C to –40°C<br />
High Temperature Exposure Test: +120°C<br />
Low Temperature Exposure Test: -30°C<br />
Standards: JIS D 0204, 0208<br />
(3)PDP environmental tests<br />
High Temperature (storage): +30 to +100°C<br />
Low Temperature (storage)*: -50 to 0°C<br />
High Temperature, High Humidity (constant): +40 to +60°C, 85% or 93%rh<br />
Low Temperature Operation: -20 to 0°C<br />
(*The Ultra View Temperature (& Humidity) Chamber is compatible down to<br />
-40°C.)<br />
To meet the demand for these standards, Espec has utilized glass reinforced with deposit<br />
metal, broadly extending the temperature range from –40°C to +120°C.<br />
The new chambers offer a greatly extended range of control for humidity as well as for<br />
temperature.<br />
- 24 -<br />
Espec Technology Report No20