STEP 6:<br />

INSTALLATION AND CALIBRATION<br />

Adjust R22 (R21, R20) to give a reading of 1300.0 oC for PtRh 6%,13% and 10% and<br />

1100.0 oC for Platinel II and type K.<br />

STEP 7:<br />

Repeat steps 5 and 6 for zone 2 and 3, adjusting R21 and R20, respectively, in step 6.<br />

<br />

NOTE<br />

Steps 1 to 4 of the above mentioned calibration procedure should be<br />

repeated for every cold junction board present.<br />