STEP 6:<br />
INSTALLATION AND CALIBRATION<br />
Adjust R22 (R21, R20) to give a reading of 1300.0 oC for PtRh 6%,13% and 10% and<br />
1100.0 oC for Platinel II and type K.<br />
STEP 7:<br />
Repeat steps 5 and 6 for zone 2 and 3, adjusting R21 and R20, respectively, in step 6.<br />
<br />
NOTE<br />
Steps 1 to 4 of the above mentioned calibration procedure should be<br />
repeated for every cold junction board present.<br />