The R 2 value was used to assess the accuracy of each concentration measurement <strong>from</strong> this<br />

project. Table 5-3, taken <strong>from</strong> the Boreal Laser, Inc. <strong>Gas</strong>Finder 2.0 Operation Manual, presents<br />

a range of R 2 values, and the corresponding accuracy of the measurement.<br />

Table 5-3. Accuracy of Concentration Measurements for Different R 2 Value<br />

R 2<br />

Measurement<br />

Accuracy (%)<br />

> 0.95 ± 2<br />

0.9 ± 5<br />

0.7 ± 10<br />

0.5 ± 15<br />

0.4 ± 20<br />

0.3 ± 25<br />

0.15 ± 50<br />

0.1 ± 70<br />

< 0.05 ± 100<br />

The R 2 value of each data point (measured methane concentration) was analyzed to assess<br />

whether or not it met the DQI criterion for accuracy of ±20 percent, which corresponds to an R 2<br />

value of greater than 0.4. A total of 21,467 data points were analyzed, and 16,509 met the DQI<br />

criteria for accuracy, for a total completeness of 77 percent. This value did not meet the project<br />

DQI criteria of 80-percent completeness.<br />

The precision of the OP-TDLAS methane concentration measurements was assessed by<br />

comparing consecutive methane concentration values measured along the same beam path.<br />

Methane path-averaged concentration data values were collected with the OP-TDLAS instrument<br />

and input into the VRPM algorithm<br />

5.2.2 DQI Check for Analyte PIC Measurement with OP-FTIR<br />

The precision and accuracy of the analyte path-integrated concentration (PIC) measurements<br />

collected with the OP-FTIR instrument was assessed by analyzing the measured nitrous oxide<br />

concentrations in the atmosphere. A typical background atmospheric concentration for nitrous<br />

oxide is about 315 ppb. However, this value may fluctuate due to seasonal variations in nitrous<br />

oxide concentrations or elevation of the site.<br />

The precision of the analyte PIC measurements was evaluated by calculating the relative<br />

standard deviation of each data subset. A subset is defined as the data collected along one<br />

particular path length during one particular survey in one survey sub-area. The number of data<br />

points in a data subset depends on the number of cycles used in a particular survey.<br />

The accuracy of the analyte PIC measurements was evaluated by comparing the calculated<br />

nitrous oxide concentrations <strong>from</strong> each data subsets to the typical background concentration of<br />