5.1 Equipment Calibration<br />

Chapter 5<br />

Quality Assurance/Quality Control<br />

All project instrumentation is calibrated annually or verified as part of standard operating<br />

procedures. Certificates of calibration are kept on file. Maintenance records are kept for any<br />

equipment adjustments or repairs in bound project notebooks that include the data and<br />

description of maintenance performed. Instrument calibration procedures and frequency are<br />

listed in Table 5-1 and further described in the text.<br />

Table 5-1. Instrumentation Calibration Frequency and Description<br />

Instrument Measurement Calibration Date Calibration Detail<br />

Boreal Methane <strong>Gas</strong>Finder 2.0 Methane PIC Pre-deployment and in-field Reference cell calibration<br />

OP-TDLAS checks<br />

R.M. Young Wind Speed in meters per 7 June 2006 APPCD Metrology Lab Cal. Records on file<br />

Meteorological Head second<br />

R.M. Young Wind direction in degrees 14 July 2006 APPCD Metrology Lab Cal. Records on file<br />

Meteorological Head <strong>from</strong> North<br />

Lumex 915+ Mercury Analyzer Elemental Mercury Pre-deployment and in-field Insertion of test cell<br />

Concentration checks<br />

Landtec GEM2000+ <strong>Landfill</strong> <strong>Gas</strong> O2, CO2, CH4, CO, H2S, Pre-deployment and in-field Calibrated at rental location before testing<br />

Monitor and balance N2 checks and in-field checks using calibration gas<br />

cylinder<br />

Shortridge ADM-870 Airdata Header pipe gas velocity 8 July 2007 Yearly factory certified NIST traceable<br />

Multimeter calibration<br />

Environmental Supply VOST <strong>Gas</strong> sample volume for 3 October 2007 EPA Method 2A<br />

Meter Box sorbent tubes<br />

Topcon Model GTS-211D Distance Measurement 19 April 2006 Calibration of distance measurement.<br />

Theodolite Actual distance=19.6 m<br />

#1 Measured distance= 19.56 m<br />

#2 Measured distance= 19.55 m<br />

Topcon Model GTS-211D Angle Measurement 19 April 2006 Calibration of angle measurement.<br />

Theodolite Actual angle= 360º<br />

#1 Measured angle= 360º28’47”<br />

#2 Measured angle= 359º39’24”<br />

5-1