An Investigation Into Ball Grid Array Inspection Techniques
An Investigation Into Ball Grid Array Inspection Techniques An Investigation Into Ball Grid Array Inspection Techniques
Oesa STM-261 5.4 Comments On Sample 2 Metallographic sectioning confirmed the bridge at P2/P3, missing balls at R I, K8 and R8 and non-wetted pads at E I0, L4 and K12. Lower volume joints were also noted at PI and R2. These may have been caused by the formation of a larger bridge around P2/P3 during reflow, with the P2/P3 bridge dominating, drawing solder away from the joints at P I and R2. Not all defects on this sample were examined metallographically. The bridges or shorts were easily identified by all X-ray inspection systems. Similarly, missing balls were identified by all systems but were more easily identified by the higher anode voltage systems. With a higher anode voltage, the low density images created by the reflowed solder paste on the substrate, are lighter showing the missing balls more easily. The non-wetted pads were less easily located by perpendicular transmission X-ray systems. These defects show themselves as lacking the tear-drop shape and also may be slightly out of normal alignment. Identification of non-wetted pads would be difficult without the tear-drop pad. Careful analysis of the stored images allows identification of these defects but some were not identified during real-time analysis. Undoubtedly, detection of non-wetted pads will be more successful as operators become more experienced. The angled transmission X-ray system showed the non-wetted pads with much greater ease. The poor wetting angle to the pad is clearly shown for example in Figure 37. All non-wetted pads were found during the real-time inspection. The laminography system automatically detected the bridges around Ll and P3. The missing ball at H8 was classified as an insufficient joint, whilst the other three missing balls were identified as missing. The non-wetted pads at EIO, KI2 and L4 were classified as misaligned. Additionally, voiding in excess of 10% of joint area was found at joints at A7, All, 88, 810, C5, E15, F14, G2, H2, K13, L3, L6, L9, LIO, M8, PIO and P 12. This threshold can be altered by the user. Thirty two apparent false detects were also registered. Hewlett Packard claim that these could be 'tuned out' if a larger production volume and more programming time were available. It should also be noted that as these samples were assembled using rework processes, the joint volume may not be as uniform as high volume production samples and this could account for some of the false detects. It should be noted that this investigation was based around one off, unique test samples, whereas, the 5DX is designed for rapid in line testing of many similar samples. 34
5.5 Results For Sample 3 The defects contained on sample 3 are: Missing balls: Bridges: Non-wetted pads: C5, F2 PI5 / RI5 PIO, C13, F13, MS, G9, K9 Figure Machine used Defects Figure Summary 3. 44 Glenbrook Bridge and non-wetted pads (K9, M8, P10) 45 Nicolet Missing (F2, C5), non-wetted C13, F13, G9, K9, M8, P10 and bridge P15 46 X-tek perpend. Missing (F2, C5), non-wetted C13, F13, G9, K9, M8,P10 and bridge P15 47 X-tek angled Bridge (P15/R15), non-wetted (M8 and P10) and solder balls (N6 + others) 48 Reichert Bridge (P15/R15) 49 Reichert Bridge (P15/R15) 50 Reichert Good joint (M15) 51 X-tek perpend. Non-wetted pads C13 and F13 52 X-tek angled Non-wetted pad (C13) and solder balls 53 Glenbrook Non-wetted pads (C13, F13 and G9) 54 Nicolet Missing balls (C5 and F2), non-wetted pads (C13, F13, G9, K9, P10) and bridge (P15) 55 Reichert Non-wetted pad (C13) 56 X-tee angled Non-wetted pad (F13) 57 HP 50X Non-wetted pads (C13, F13 and G9) 58 Reichert Non-wetted pad (F13) 59 Reichert Non-wetted pad (F13) 60 Reichert Solder balls (R15) 35 Ball Grid Array Inspection Techniques
- Page 1 and 2: An Investigation Into Ball Grid Arr
- Page 3: Abstract An evaluation of four diff
- Page 7 and 8: 1 Introduction X-ray techniques hav
- Page 9 and 10: X-rays and displays them as an imag
- Page 11 and 12: Each defect was identified using th
- Page 13 and 14: 2.3 BGA PCB Image Summary of defect
- Page 15 and 16: Sample 4 Misaligned component 2.4 U
- Page 17 and 18: 4. Equipment Evaluated Four machine
- Page 19 and 20: 'If ~ c.S"'
- Page 21 and 22: 5. Results This section is structur
- Page 23 and 24: ...~,.--= o't:ll.JOII_s_8I:Jt11l .
- Page 25 and 26: 19 Ball Grid Array Inspection Techn
- Page 27 and 28: .-:;,... ~ "''' . ..~ , '. ., ' -,
- Page 29 and 30: 5.3 Results For Sample 2 The defect
- Page 31 and 32: 25 Ball Grid Array Inspection Techn
- Page 33 and 34: 29 Ball Grid Array Inspection Techn
- Page 35 and 36: ~e 88el, e~.~tti '88.~_",! 88M_llle
- Page 37: 33 Ball Grid Array Inspection Techn
- Page 41 and 42: ~. ij.ij, --- - -- - # . '\>. ~ l 3
- Page 43 and 44: ~,-a;_:,t. ,"""~.. '~::-~-",;;:-:~,
- Page 45 and 46: . ..~.' '...~ .' _.I:t!IIDJ:I ""'1:
- Page 47 and 48: The laminography system also behave
- Page 49 and 50: 47 Ball Grid Array Inspection Techn
- Page 51 and 52: f '~"~>-" . 1 ~"I ,} ""'"''-'''''''
- Page 53 and 54: 51 Ball Grid Array Inspection Techn
- Page 55 and 56: 5.11. Identification of voids withi
- Page 57 and 58: Table 1: Defect Detectability and L
- Page 59 and 60: 7. Recommendations The most sensiti
- Page 61 and 62: 9. References 1. ESA PSS 01-708 The
- Page 63: ~esa STM-261 Editor: Copyright: Pub
Oesa STM-261<br />
5.4 Comments On Sample 2<br />
Metallographic sectioning confirmed the bridge at P2/P3, missing balls at R I, K8 and<br />
R8 and non-wetted pads at E I0, L4 and K12. Lower volume joints were also noted at PI<br />
and R2. These may have been caused by the formation of a larger bridge around P2/P3<br />
during reflow, with the P2/P3 bridge dominating, drawing solder away from the joints at<br />
P I and R2. Not all defects on this sample were examined metallographically.<br />
The bridges or shorts were easily identified by all X-ray inspection systems. Similarly,<br />
missing balls were identified by all systems but were more easily identified by the<br />
higher anode voltage systems. With a higher anode voltage, the low density images<br />
created by the reflowed solder paste on the substrate, are lighter showing the missing<br />
balls more easily.<br />
The non-wetted pads were less easily located by perpendicular transmission X-ray<br />
systems. These defects show themselves as lacking the tear-drop shape and also may be<br />
slightly out of normal alignment. Identification of non-wetted pads would be difficult<br />
without the tear-drop pad. Careful analysis of the stored images allows identification of<br />
these defects but some were not identified during real-time analysis. Undoubtedly,<br />
detection of non-wetted pads will be more successful as operators become more<br />
experienced.<br />
The angled transmission X-ray system showed the non-wetted pads with much greater<br />
ease. The poor wetting angle to the pad is clearly shown for example in Figure 37. All<br />
non-wetted pads were found during the real-time inspection.<br />
The laminography system automatically detected the bridges around Ll and P3. The<br />
missing ball at H8 was classified as an insufficient joint, whilst the other three missing<br />
balls were identified as missing. The non-wetted pads at EIO, KI2 and L4 were<br />
classified as misaligned. Additionally, voiding in excess of 10% of joint area was found<br />
at joints at A7, All, 88, 810, C5, E15, F14, G2, H2, K13, L3, L6, L9, LIO, M8, PIO<br />
and P 12. This threshold can be altered by the user. Thirty two apparent false detects<br />
were also registered. Hewlett Packard claim that these could be 'tuned out' if a larger<br />
production volume and more programming time were available. It should also be noted<br />
that as these samples were assembled using rework processes, the joint volume may not<br />
be as uniform as high volume production samples and this could account for some of the<br />
false detects. It should be noted that this investigation was based around one off, unique<br />
test samples, whereas, the 5DX is designed for rapid in line testing of many similar<br />
samples.<br />
34