07.08.2013 Views

Optoelectronics with Carbon Nanotubes

Optoelectronics with Carbon Nanotubes

Optoelectronics with Carbon Nanotubes

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

would expect the width to be comparable to kBT if the lattice temperature had a significant effect<br />

71, 73,<br />

on the carriers. On the other hand, electroluminescence typically shows much larger widths<br />

76, 85, 87, 97, 109, 111, 117 , usually a factor of few larger than PL. When FWHM vs. power is fit <strong>with</strong> a<br />

linear equation and extrapolated to zero intensity (see Figure III-8), the minimum FWHM at zero<br />

power is almost 7 times wider than 25 meV, i.e., 170 meV for this device. Other devices also<br />

show a minimum width typically between 130 and 200 meV. Although Freitag et al. attribute<br />

the fivefold EL broadening compared to PL in their data to high-temperature electrons and<br />

phonons, they do not relate the broadening to the temperatures quantitatively. There are actually<br />

several possible contributions to the width, such as tube heterogeneity, phonon scattering,<br />

external electric field, exciton-exciton annihilation, electronic temperature, and blackbody<br />

background, just to name a few. Some of these contributions are now discussed semi-<br />

quantitatively.<br />

The minimum width at zero power points to tube heterogeneity as one of the main<br />

contributions to the large width. When only a single peak is used to fit the spectra, the sharpness<br />

of the main peak is not reproduced by Equation III.1 (0.71 eV in Figure III-4 (a)), and this is<br />

always the case <strong>with</strong> any of our CNTFET EL spectra. This suggests that the “main peak” is<br />

actually a combination of multiple peaks that are broadened and combined to form one very<br />

broad shape, obscuring the contribution by each peak. By examining Figure III-4 (a) more<br />

closely, one can observe at least three peaks, at 0.66 eV, 0.71 eV and 0.88 eV. While the 0.66<br />

eV is most easily seen at VDS = -7.0 eV, it gets obscured somewhat as |VDS| is increased. From<br />

observations of double peaks in p-n junctions (see Chapter IV), this lower-energy peak is<br />

considered to be a weakly localized excitonic peak, possibly bound to local defect sites. This<br />

double-peak feature is discussed in greater detail in the section on single-tube p-n junction diodes<br />

where the peaks are observed more distinctly, but suffice it to say that this alone adds 50 meV<br />

(i.e., the difference between the peak positions) to the total width, provided that the two peaks<br />

are of comparable intensity. The energy difference was even higher in some devices, up to 65<br />

meV.<br />

It also needs to be noted that these peaks are unlikely due to the aggregation of multiple<br />

tubes in a bundle. As explained in the Methods section, bulk tubes (which are originally<br />

bundled) are sonicated for at least 30 minutes at a high power in dichloroethane before deposited<br />

onto the substrate, which is sufficient to separate them into individual tubes, as we have rarely<br />

49

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!