<strong>Disposable</strong> <strong>sensors</strong> <strong>that</strong> <strong>last</strong><br />

Durable and cost effective single patient use SpO 2 <strong>sensors</strong><br />

<strong>Philips</strong> family of single-patient use SpO2 <strong>sensors</strong> provides continuous, non-invasive<br />

measurement of arterial oxygen saturation.<br />

You can choose between our Adult-Pediatric<br />

sensor to measure a wide range of patients or<br />

our wrap <strong>sensors</strong> <strong>that</strong> are optimally designed<br />

for the infant or neonate.<br />

In addition, these <strong>sensors</strong> are optimized for<br />

<strong>Philips</strong> FAST-SpO technology, providing you<br />

2<br />

additional benefits such as reduced sensitivity<br />

to patient movement and other artifacts.<br />

When your goal is superb patient care,<br />

there’s no better choice than <strong>Philips</strong>.<br />

<strong>Philips</strong> family of disposable <strong>sensors</strong><br />

• Cost-effective. <strong>Philips</strong> <strong>Disposable</strong>s offer a<br />

cost-effective solution and when combined<br />

with reusable <strong>sensors</strong>, can help to lower<br />

your total cost of SpO monitoring.<br />

2<br />

• Quality. <strong>Philips</strong> meets a wide range of<br />

standards <strong>that</strong> have been set by regulatory<br />

authorities and standardization committees<br />

(e.g. FDA, IEC, ISO), plus <strong>Philips</strong> own<br />

internal standards to achieve high levels of<br />

quality and safety.<br />

• Compatible. Tested and qualified<br />

for <strong>Philips</strong>, legacy HP and Agilent, and<br />

validated Nellcor ® and GE monitors.<br />

• Convenient. 20 <strong>sensors</strong> per box.


Three <strong>sensors</strong> for a wide range of patient sizes<br />

M1131A (Adult/Pediatric)<br />

• Versatile. One sensor fits pediatric to<br />

large adult patients.<br />

• Simple to use. Light ergonomically<br />

shaped sensor slips easily onto a patient’s<br />

finger and is easy to reposition as needed.<br />

• Reliable. Pre-aligned LED sensor provides<br />

fast, reliable readings you can count on to<br />

make informed clinical decisions.<br />

• Long <strong>last</strong>ing. Designed to perform for<br />

a single patient’s average hospital stay.<br />

M1132A and M1133A<br />

(Infant and Neonate)<br />

• Comfortable. Minimal adhesive is<br />

optimized to reduce irritation when<br />

skin sensitivity is a concern.<br />

• Accurate. Sensor design prevents ambient<br />

light for reliable readings.<br />

• Durable and cost effective.<br />

Ruggedly designed to out<strong>last</strong> typical adhesive<br />

<strong>sensors</strong>.<br />

© 2008 Koninklijke <strong>Philips</strong> Electronics N.V.<br />

All rights are reserved.<br />

M1131A (Adult/Pediatric Sensor)<br />

• Patient size: greater than 20kg (44 lb)<br />

• Application: any finger (except thumb)<br />

M1132A (Infant Sensor)<br />

• Patient size: 3-10kg (6.6-22 lb)<br />

• Application: any finger or toe<br />

M1133A (Neonatal/Infant/Adult<br />

Sensor)<br />

• Patient size: 40kg (>88 lb)<br />

Application: finger<br />

* Pulse oximetry measurements are statistically distributed. Two-thirds of all pulse oximetry measurements can be expected to fall within the stated accuracy.<br />

When <strong>sensors</strong> are used on neonatal subjects as recommended, the literature predicts the accuracy range should be increased by +/-1% to account for affect<br />

on oximeter measurements due to fetal hemoglobin in neonatal blood.<br />

<strong>Philips</strong> Medical Systems Nederland B.V. reserves the right to make changes<br />

in specifications and/or to discontinue any product at any time without<br />

notice or obligation and will not be liable for any consequences resulting<br />

from the use of this publication.<br />

<strong>Philips</strong> Healthcare is part of Royal <strong>Philips</strong> Electronics<br />

www.medical.philips.com<br />

medical@philips.com<br />

fax: +31 40 27 64 887<br />

Specifications<br />

Physical<br />

M1131A:<br />

Cable length: 45.0 cm (18”)<br />

M1132A, M1133A:<br />

Cable length: 90.0 cm (35”)<br />

Temperature range<br />

Operating: 10 to 37°C<br />

(50 to 98.6°F)<br />

Humidity<br />

Operating: 95% RH max at<br />

37°C (98.6°F)<br />

SpO 2 accuracy*<br />

Accuracy with compatible<br />

<strong>Philips</strong> (Agilent/Hewlett-Packard)<br />

and Nellcor monitors: (N-200,<br />

N-395, N3000 N-20 PA, NPB-40)<br />

and GE Dinamap Pro 400V1.<br />

M1131A:<br />

RMS = ±3% (70 to 100%)<br />

M1132A, M1133A:<br />

RMS = ±2% (70 to 100%)<br />

Printed in The Netherlands<br />

4522 962 37351 *NOV 2008<br />

<strong>Philips</strong> Healthcare<br />

Global Information Center<br />

P.O. Box 1286<br />

5602 BG Eindhoven<br />

The Netherlands