<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 66 of 82<br />

Table 71: Attribute Mapping of the Veradius AE<br />

Attribute Name MWL Tag<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009<br />

MPPS<br />

Create Tag Set Tag<br />

SC Tag XA Tag<br />

Specific Character Set (if present) 0008,0005 0008,0005 - 0008,0005 0008,0005<br />

Accession Number 0008,0050 0008,0050 - 0008,0050 0008,0050<br />

Modality 0008,0060 0008,0060 - 0008,0060 0008,0060<br />

Referring Physician’s Name 0008,0090 - - 0008,0090 0008,0090<br />

Referenced Study Sequence 0008,1110 0008,1110 - 0008,1110 0008,1110<br />

> Referenced SOP Class UID 0008,1150 0008,1150 - - 0008,1150<br />

> Referenced SOP Instance UID 0008,1155 0008,1155 - - 0008,1155<br />

Referenced Patient Sequence 0008,1120 0008,1120 - - -<br />

> Referenced SOP Class UID 0008,1150 0008,1150 - - -<br />

> Referenced SOP Instance UID 0008,1155 0008,1155 - - -<br />

Patient's Name 0010,0010 0010,0010 - 0010,0010 0010,0010<br />

Patient ID 0010,0020 0010,0020 - 0010,0020 0010,0020<br />

Patient's Birth Date 0010,0030 0010,0030 - 0010,0030 0010,0030<br />

Patient's Birth Time 0010,0032 - - 0010,0032 0010,0032<br />

Patient's Sex 0010,0040 0010,0040 - 0010,0040 0010,0040<br />

Other Patient IDs 0010,1000 - - 0010,1000 0010,1000<br />

Other Patient Names 0010,1001 - - 0010,1001 0010,1001<br />

Patient's Weight 0010,1030 - - 0010,1030 0010,1030<br />

Study Instance UID 0020,000D 0020,000D - 0020,000D 0020,000D<br />

Requested Procedure Description 0032,1060 0032,1060 - - -<br />

Scheduled Performing Physician’s<br />

Name (Physician who makes the<br />

Examination)<br />

0040,0006 - 0008,1050 0008,1050 0008,1050<br />

Scheduled Procedure Step Description 0040,0007 0040,0007 - - -<br />

Scheduled Procedure Step ID 0040,0009 0040,0009 - - -<br />

> Scheduled Performing Physician's<br />

Name<br />

0040,0006 - 0008,1050 0008,1050 0008,1050<br />

> Scheduled Procedure Step<br />

Description<br />

0040,0007 0040,0007 - - -<br />

> Scheduled Protocol Code Sequence 0040,0008 0040,0008 - - -<br />

>> Code Value 0008,0100 0008,0100 - - -<br />

>> Coding Scheme Designator 0008,0102 0008,0102 - - -<br />

>> Coding Scheme Version 0008,0103 0008,0103 - - -<br />

>> Code Meaning 0008,0104 0008,0104 - - -<br />

> Scheduled Procedure Step ID 0040,0009 0040,0009 - - -<br />

Requested Procedure ID 0040,1001 0040,1001 - - -<br />

8.1.4. Coerced/Modified fields<br />

When exporting an image the following behavior applies.<br />

A Secondary Capture image shall be exported as reflected in the GUI.<br />

To enable reconstruction, an X-ray image shall be exported without annotations and<br />

using the original grayscale values as per acquisition.