<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 64 of 82<br />

Name Tag VR<br />

Present<br />

of Value<br />

Source Comment<br />

General Equipment Module<br />

Manufacturer 0008,0070 LO ALWAYS AUTO Applied<br />

value:<br />

<strong>Philips</strong><br />

Medical<br />

Systems<br />

Institution Name 0008,0080 LO ALWAYS CONFIG Hospital<br />

Name.<br />

Station Name 0008,1010 SH ALWAYS CONFIG -<br />

Manufacturer's Model Name 0008,1090 LO ALWAYS AUTO Applied<br />

value:<br />

Veradius<br />

General Image Module<br />

Content Date 0008,0023 DA ALWAYS AUTO -<br />

Content Time 0008,0033 TM ALWAYS AUTO -<br />

Instance Number 0020,0013 IS ALWAYS AUTO -<br />

Patient Orientation 0020,0020 CS EMPTY FIXED -<br />

Image Pixel Module<br />

Rows 0028,0010 US ALWAYS AUTO Applied<br />

value: 1024<br />

Columns 0028,0011 US ALWAYS AUTO Applied<br />

value: 1024<br />

Pixel Data 7FE0,0010 OW ALWAYS AUTO -<br />

Cine Module<br />

Start Trim 0008,2142 IS ALWAYS AUTO Applied<br />

value: 1<br />

Stop Trim 0008,2143 IS ALWAYS AUTO Number of<br />

images in<br />

the run.<br />

Recommended Display Frame Rate 0008,2144 IS ANAP AUTO Acquisition<br />

speed.<br />

Cine Rate 0018,0040 IS ANAP AUTO Calculated<br />

from<br />

acquisition<br />

speed.<br />

Frame Time 0018,1063 DS ALWAYS AUTO Calculated<br />

from<br />

acquisition<br />

speed [ms].<br />

Multi-Frame Module<br />

Number of Frames 0028,0008 IS ALWAYS AUTO Number of<br />

exported<br />

images in the<br />

run.<br />

Frame Increment Pointer 0028,0009 AT ALWAYS AUTO Applied<br />

value:<br />

0x00181063<br />

(Frame Time)<br />

X-Ray Image Module<br />

Image Type 0008,0008 CS ALWAYS AUTO Applied<br />

value:<br />

ORIGINAL\<br />

PRIMARY<br />

Samples per Pixel 0028,0002 US ALWAYS AUTO Applied<br />

value: 1<br />

Photometric Interpretation 0028,0004 CS ALWAYS AUTO Applied<br />

value:<br />

MONOCHR<br />

OME2<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009