03.08.2013 Views

DICOM Conformance Statement - InCenter - Philips

DICOM Conformance Statement - InCenter - Philips

DICOM Conformance Statement - InCenter - Philips

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 60 of 82<br />

Information Entity Module Name Usage<br />

SOP Common module ALWAYS<br />

Table 68: Created Secondary Capture Image Storage SOP Class Attributes<br />

Name Tag VR<br />

Present<br />

of Value<br />

Source Comment<br />

Patient Module<br />

Patient's Name 0010,0010 PN ALWAYS MWL /<br />

USER<br />

-<br />

Patient ID 0010,0020 LO ALWAYS MWL /<br />

USER<br />

-<br />

Patient's Birth Date 0010,0030 DA ALWAYS MWL /<br />

USER<br />

-<br />

Patient's Birth Time 0010,0032 TM VNAP MWL -<br />

Patient's Sex 0010,0040 CS ALWAYS MWL /<br />

USER<br />

-<br />

Other Patient IDs 0010,1000 LO VNAP MWL -<br />

Other Patient Names 0010,1001 PN VNAP MWL -<br />

General Study Module<br />

Study Date 0008,0020 DA ALWAYS AUTO Examination<br />

date.<br />

Study Time 0008,0030 TM ALWAYS AUTO Examination<br />

time.<br />

Accession Number 0008,0050 SH ALWAYS MWL /<br />

USER<br />

-<br />

Referring Physician's Name 0008,0090 PN VNAP MWL -<br />

Study Description 0008,1030 LO ALWAYS IMPLICIT<br />

/ MWL<br />

Procedure Code Sequence 0008,1032 SQ ANAP MWL -<br />

Referenced Study Sequence 0008,1110 SQ VNAP MWL -<br />

> Referenced SOP Class UID 0008,1150 UI ALWAYS MWL -<br />

> Referenced SOP Instance UID 0008,1155 UI ALWAYS MWL -<br />

Study Instance UID 0020,000D UI ALWAYS AUTO /<br />

MWL<br />

-<br />

Study ID 0020,0010 SH EMPTY FIXED -<br />

Patient Study Module<br />

Patient’s Weight 0010,1030 DS VNAP MWL -<br />

General Series Module<br />

Performing Physician’s Name 0008,1050 PN VNAP,<br />

USER<br />

User selected<br />

examination<br />

type or copied<br />

from requested<br />

procedure<br />

description<br />

(0032,1060)<br />

received via<br />

WLM.<br />

Can be entered<br />

by Physician<br />

who makes the<br />

Examination<br />

Series Instance UID 0020,000E UI ALWAYS AUTO -<br />

Series Number 0020,0011 IS ALWAYS AUTO Increasing<br />

number that<br />

identifies series<br />

(run).<br />

Laterality 0020,0060 CS EMPTY FIXED -<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009