7. SECURITY<br />

7.1. Security Profiles<br />

<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 57 of 82<br />

7.1.1. Basic Application Level Confidentiality Profile<br />

The Veradius AE conforms to the Basic Application Level Confidentiality Profile as deidentifier.<br />

De-identified SOP Instances will be created on <strong>DICOM</strong> Media if specified by the user.<br />

No instances of the Encrypted Attributes Data Set are created. No transfer syntaxes<br />

are supported for encoding/decoding of Encrypted Attributes Data Sets.<br />

The terms used to describe the replacement value in the anonymized patient data can<br />

be read as follows:<br />

COPY Same value as in source data<br />

EMPTY The attribute will have a value of zero length.<br />

ANP Attribute Not Present<br />

n.a. Not applicable, the attribute is not contained in the standard IOD of the<br />

Veradius AE<br />

The next table lists the protected data attributes.<br />

Table 66: Basic Application Level Confidentiality Profile Attributes<br />

Name Tag VR Replacement Value<br />

Instance Creator UID 0008,0014 UI n.a.<br />

SOP Instance UID 0008,0018 UI COPY<br />

Accession Number 0008,0050 SH EMPTY<br />

Institution Name 0008,0080 LO ANP<br />

Institution Address 0008,0081 ST n.a.<br />

Referring Physician’s Name 0008,0090 PN EMPTY<br />

Referring Physician’s Address 0008,0092 ST n.a.<br />

Referring Physician’s Telephone Numbers 0008,0094 SH n.a.<br />

Station Name 0008,1010 SH COPY<br />

Study Description 0008,1030 LO COPY<br />

Series Description 0008,103E LO COPY<br />

Institutional Department Name 0008,1040 LO n.a.<br />

Physician(s) of Record 0008,1048 PN n.a.<br />

Performing Physicians’ Name 0008,1050 PN ANP<br />

Name of Physician(s) Reading Study 0008,1060 PN n.a.<br />

Operators’ Name (Technologist) 0008,1070 PN COPY<br />

Admitting Diagnoses Description 0008,1080 LO n.a.<br />

Referenced SOP Instance UID 0008,1155 UI COPY<br />

Derivation Description 0008,2111 ST COPY<br />

Patient’s Name 0010,0010 PN EMPTY<br />

Patient ID 0010,0020 LO In Patient Module the Patient ID<br />

value is “EMPTY”.<br />

In the DIRECTORY RECORD: 0<br />

(PATIENT) the Patient ID value<br />

has a new generated value<br />

Patient’s Birth Date 0010,0030 DA EMPTY<br />

Patient’s Birth Time 0010,0032 TM COPY<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009