DICOM Conformance Statement - InCenter - Philips
DICOM Conformance Statement - InCenter - Philips
DICOM Conformance Statement - InCenter - Philips
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
7. SECURITY<br />
7.1. Security Profiles<br />
<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 57 of 82<br />
7.1.1. Basic Application Level Confidentiality Profile<br />
The Veradius AE conforms to the Basic Application Level Confidentiality Profile as deidentifier.<br />
De-identified SOP Instances will be created on <strong>DICOM</strong> Media if specified by the user.<br />
No instances of the Encrypted Attributes Data Set are created. No transfer syntaxes<br />
are supported for encoding/decoding of Encrypted Attributes Data Sets.<br />
The terms used to describe the replacement value in the anonymized patient data can<br />
be read as follows:<br />
COPY Same value as in source data<br />
EMPTY The attribute will have a value of zero length.<br />
ANP Attribute Not Present<br />
n.a. Not applicable, the attribute is not contained in the standard IOD of the<br />
Veradius AE<br />
The next table lists the protected data attributes.<br />
Table 66: Basic Application Level Confidentiality Profile Attributes<br />
Name Tag VR Replacement Value<br />
Instance Creator UID 0008,0014 UI n.a.<br />
SOP Instance UID 0008,0018 UI COPY<br />
Accession Number 0008,0050 SH EMPTY<br />
Institution Name 0008,0080 LO ANP<br />
Institution Address 0008,0081 ST n.a.<br />
Referring Physician’s Name 0008,0090 PN EMPTY<br />
Referring Physician’s Address 0008,0092 ST n.a.<br />
Referring Physician’s Telephone Numbers 0008,0094 SH n.a.<br />
Station Name 0008,1010 SH COPY<br />
Study Description 0008,1030 LO COPY<br />
Series Description 0008,103E LO COPY<br />
Institutional Department Name 0008,1040 LO n.a.<br />
Physician(s) of Record 0008,1048 PN n.a.<br />
Performing Physicians’ Name 0008,1050 PN ANP<br />
Name of Physician(s) Reading Study 0008,1060 PN n.a.<br />
Operators’ Name (Technologist) 0008,1070 PN COPY<br />
Admitting Diagnoses Description 0008,1080 LO n.a.<br />
Referenced SOP Instance UID 0008,1155 UI COPY<br />
Derivation Description 0008,2111 ST COPY<br />
Patient’s Name 0010,0010 PN EMPTY<br />
Patient ID 0010,0020 LO In Patient Module the Patient ID<br />
value is “EMPTY”.<br />
In the DIRECTORY RECORD: 0<br />
(PATIENT) the Patient ID value<br />
has a new generated value<br />
Patient’s Birth Date 0010,0030 DA EMPTY<br />
Patient’s Birth Time 0010,0032 TM COPY<br />
© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />
XPS031-090132.02 Veradius R1.1 6 Jul 2009