<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 53 of 82<br />
5.2.1.2.2.1. Media Storage Application Profile<br />
As depicted in Table 61, the VF Workstation AE supports the RWA Read Image for<br />
STD-GEN-CD, STD-GEN-USB-JPEG and STD-GEN-DVD-JPEG application profiles.<br />
5.2.1.2.2.1.1. Options<br />
The mandatory attributes of the <strong>DICOM</strong> images are required for the correct storage of<br />
the images in the local database. Optional attributes and retired/private attributes are<br />
stored too – if present; this is equivalent with the level 2 (Full) conformance for the<br />
Storage service class in the Network support.<br />
5.2.1.2.3. Write Image<br />
When an image transfer to <strong>DICOM</strong> media is initiated then the VF Workstation AE acts<br />
as an FSC using the interchange option to write SOP instances on the <strong>DICOM</strong> media.<br />
5.2.1.2.3.1. Media Storage Application Profile<br />
As depicted in Table 61, the VF Workstation AE supports the RWA Write Image for<br />
STD-GEN-CD, STD-GEN-USB-JPEG and STD-GEN-DVD-JPEG application profiles.<br />
However, the VF Workstation AE only supports writing on DVD+R(W) media, not<br />
DVD-R(W) media.<br />
5.2.1.2.3.1.1. Options<br />
The <strong>DICOM</strong>DIR file will be extended when new images are written. In case some<br />
attributes are not present in an image but are specified as mandatory in the<br />
<strong>DICOM</strong>DIR definition of <strong>DICOM</strong> media, a generated value will be filled in.<br />
Implementation remarks an restrictions<br />
When writing the <strong>DICOM</strong>DIR records, key values are generated when no value of the<br />
corresponding attribute is supplied, according to the following table.<br />
Table 63: Generated Keys<br />
Key Tag Generated Value<br />
Patient Keys<br />
Patient ID 0010,0020 At import the VF Workstation AE each time creates a new value based<br />
on the Study Instance UID for each new study written to <strong>DICOM</strong> media<br />
(even if this study belongs to a patient recorded earlier).<br />
Otherwise the default generated value shall be a succession of<br />
“UNKNOWN”, the Patient’s Name, the Patient’s Birth Date, and the<br />
Patient’s Sex, concatenated by using underscore characters.<br />
Study Keys<br />
Study Date 0008,0020 Current date.<br />
Study Time 0008,0030 Current time.<br />
Study ID 0020,0010 “UNKNOWN”<br />
Series Number 0020,0011 1<br />
Instance Number 0020,0013 1<br />
Series Keys<br />
Image Keys<br />
The default value for (0028,1040) Pixel Intensity Relationship is set to DISP.<br />
The VF Workstation AE can write volumes of the media to that media.<br />
If spanning is required then the VF Workstation AE asks for a new media.<br />
© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />
XPS031-090132.02 Veradius R1.1 6 Jul 2009