<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 41 of 82<br />

Figure 12: Sequencing of RWA Query/Retrieve Image<br />

For each retrieve request (selected from query results) the VF Workstation AE<br />

accepts an association from the selected peer entity (Archive) and uses it to receive<br />

image Storage (C-STORE) requests and send subsequent responses. On request of<br />

the Storage SCU (Archive) the association is released.<br />

4.2.2.4.1.2. Accepted Presentation Contexts<br />

The VF Workstation AE will accept Presentation Contexts as shown in Table 49.<br />

Table 49: Acceptable Presentation Contexts for Query/Retrieve Image<br />

Presentation Context Table<br />

Abstract Syntax Transfer Syntax<br />

Name UID Name List UID List<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009<br />

Role<br />

Extended<br />

Negotiation<br />

Any defined SOP class ELE 1.2.840.10008.1.2.1 SCP None<br />

EBE 1.2.840.10008.1.2.2<br />

ILE 1.2.840.10008.1.2<br />

Note: For performance reasons the ELE transfer syntax is preferred and shall be chosen in case<br />

multiple transfer syntaxes are proposed in the association negotiation.<br />

The VF Workstation AE shall accept all contexts in the intersection of the proposed<br />

and acceptable presentation contexts. This means that the VF Workstation AE<br />

accepts multiple proposed presentation contexts with the same SOP class but<br />

different transfer syntaxes.<br />

There is no check for duplicate contexts, and these will therefore be accepted.<br />

4.2.2.4.1.3. SOP Specific <strong>Conformance</strong> for SOP Classes<br />

4.2.2.4.1.3.1. Verification<br />

The VF Workstation AE provides standard conformance to the Verification service<br />

class.<br />

The status code behavior is as summarized in Table 50.