<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 37 of 82<br />
Figure 11: Sequencing of RWA Query/Retrieve Image<br />
The operator queries a remote archive, using the query tool in the data handling<br />
facility. The VF Workstation AE initiates an association to the selected peer entity<br />
(Archive) and uses it to send Query (C-FIND) requests and receive subsequent<br />
responses. The association is released when the execution of the query completes<br />
and the Query/Retrieve dialog on the GUI is closed. The matching images are then<br />
displayed in a patient folder for the remote archive.<br />
The required images can now be selected for copying to the Veradius, using the copy<br />
tool in the data handling facility. For each copy request the VF Workstation AE<br />
initiates an association to the selected peer entity (Archive) and uses it to send<br />
Retrieve (C-MOVE) requests and receive subsequent responses; an examination may<br />
contain both images and presentation states. The association is released after the<br />
final Retrieve (C-MOVE) response for the related request has been received (no more<br />
pending).<br />
4.2.2.3.1.2. Proposed Presentation Contexts<br />
For Query/Retrieve Image the VF Workstation AE will propose the following<br />
presentation contexts.<br />
© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />
XPS031-090132.02 Veradius R1.1 6 Jul 2009