<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 34 of 82<br />

The Veradius AE will only accept the SCU role (which must be proposed via<br />

SCP/SCU Role Selection Negotiation) within a Presentation Context for the Storage<br />

Commitment Push Model SOP Class.<br />

4.2.1.4.1.3. SOP Specific <strong>Conformance</strong> for SOP Classes<br />

The behavior of the Veradius AE when receiving Event Types within the N-EVENT-<br />

REPORT is summarized in Table 37.<br />

Event Type Name<br />

Storage Commitment<br />

Request Successful<br />

Storage Commitment<br />

Request Complete –<br />

Failures Exist<br />

Table 37: Storage Commitment - N-EVENT-REPORT Behavior<br />

Event<br />

Behavior<br />

Type ID<br />

1 The Referenced SOP Instances under Referenced SOP Sequence<br />

(0008,1199) are marked within the database as “Stored & Committed (SC)”<br />

to the value of Retrieve AE Title (0008,0054).<br />

2 In case of a “Failure Exist” situation (Referenced SOP Instances under Failed<br />

SOP Sequence (0008,1198)), all of the stored SOP Instances for that<br />

examination are considered as failed for storage commitment. A send job<br />

that failed storage commitment will not be automatically restarted but can be<br />

resumed by the user.<br />

The status response behavior of the Veradius AE is as summarized in Table 38.<br />

Table 38: Storage Commitment - N-EVENT-REPORT Status Response<br />

Service Status Code Further Meaning Description<br />

Success 0000 Success The Veradius AE has completed the operation<br />

successfully.<br />

Failure * Any other Failure status<br />

code<br />

The association is aborted and the storage commit<br />

N-EVENT-REPORT is marked as failed.<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009