DICOM Conformance Statement - InCenter - Philips
DICOM Conformance Statement - InCenter - Philips
DICOM Conformance Statement - InCenter - Philips
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
4.2.1.3.3. Export<br />
<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 23 of 82<br />
4.2.1.3.3.1. Description and Sequencing of Activities<br />
After an acquisition the Veradius AE sends related MPPS data to a Study<br />
Management SCP (RIS/HIS). Then the acquired image is stored or printed according<br />
the settings as specified by the operator.<br />
Figure 8: Sequencing of RWA Export (store)<br />
The acquisition is initiated by selecting an examination for Export. After selecting<br />
“MPPS & OK” the protocol name and study status have to be selected. Then the<br />
© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />
XPS031-090132.02 Veradius R1.1 6 Jul 2009