03.08.2013 Views

DICOM Conformance Statement - InCenter - Philips

DICOM Conformance Statement - InCenter - Philips

DICOM Conformance Statement - InCenter - Philips

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 20 of 82<br />

Table 13: Matching Criteria for Identifying Worklist Entries<br />

Attribute<br />

Name Tag VR Comment<br />

Scheduled Procedure Step ID (0040,0009) SH<br />

Accession Number (0008,0050) SH<br />

Requested Procedure ID (0040,1001) SH<br />

If none of these identification attributes is present then the received worklist entry is<br />

ignored.<br />

4.2.1.3.2.2. Proposed Presentation Contexts<br />

For Get Worklist the Veradius AE will propose the following presentation contexts.<br />

Table 14: Proposed Presentation Contexts for Get Worklist<br />

Presentation Context Table<br />

Abstract Syntax Transfer Syntax<br />

Name UID Name List UID List<br />

Modality<br />

Worklist<br />

Information<br />

Model – FIND<br />

Modality<br />

Performed<br />

Procedure<br />

Step<br />

1.2.840.10008.5.1.4.31 EBE<br />

ELE<br />

ILE<br />

1.2.840.10008.3.1.2.3.3 EBE<br />

ELE<br />

ILE<br />

1.2.840.10008.1.2.2<br />

1.2.840.10008.1.2.1<br />

1.2.840.10008.1.2<br />

1.2.840.10008.1.2.2<br />

1.2.840.10008.1.2.1<br />

1.2.840.10008.1.2<br />

4.2.1.3.2.3. SOP Specific <strong>Conformance</strong> for SOP Classes<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009<br />

Role<br />

SCU None<br />

SCU None<br />

Extended<br />

Negotiation<br />

4.2.1.3.2.3.1. Modality Worklist Information Model – FIND<br />

The Veradius AE provides standard conformance to the Modality Worklist SOP class.<br />

The Veradius AE can contain a number of 100 worklist entries. If the sum of current<br />

and new worklist entries exceeds 100 then the Veradius AE will release the<br />

association immediately. The Veradius AE will show a message stating that the<br />

maximum number of examinations was reached.<br />

All details regarding the specific conformance, including response behavior to all<br />

status codes, both from an application level and communication errors are provided in<br />

Table 15.<br />

Table 15: C-FIND Response Status Handling Behavior<br />

Service Status Code Further Meaning Behavior<br />

Success 0000 Matching is complete<br />

– No final identifier is<br />

supplied<br />

Failure A700 Refused<br />

– Out of resources<br />

The association is released and the matches are stored.<br />

Processing of the matches and the association is<br />

terminated. A message appears in the GUI.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!