4.2.1.3.2. Get Worklist<br />

<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 19 of 82<br />

4.2.1.3.2.1. Description and Sequencing of Activities<br />

The Veradius AE can send a modality worklist query (C-FIND) to update the Veradius<br />

worklist.<br />

Figure 7: Sequencing of RWA Get Worklist<br />

The worklist query is initiated by selecting “Get Worklist”. Then the Veradius AE opens<br />

an association and sends a modality worklist query. The BWLM SCP (RIS/HIS)<br />

returns the applicable worklist; a response with status Pending is received for each<br />

new entry, the final response has status Success. After the final response the<br />

Veradius AE releases the association.<br />

The contents of the received worklist are compared with the contents of the previous<br />

worklist. In case there are any changes, the Veradius patient file is updated. A unique<br />

match of the following attributes identifies a worklist entry.<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

XPS031-090132.02 Veradius R1.1 6 Jul 2009