DICOM Conformance Statement - InCenter - Philips
DICOM Conformance Statement - InCenter - Philips
DICOM Conformance Statement - InCenter - Philips
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
4.2.1.3.2. Get Worklist<br />
<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 19 of 82<br />
4.2.1.3.2.1. Description and Sequencing of Activities<br />
The Veradius AE can send a modality worklist query (C-FIND) to update the Veradius<br />
worklist.<br />
Figure 7: Sequencing of RWA Get Worklist<br />
The worklist query is initiated by selecting “Get Worklist”. Then the Veradius AE opens<br />
an association and sends a modality worklist query. The BWLM SCP (RIS/HIS)<br />
returns the applicable worklist; a response with status Pending is received for each<br />
new entry, the final response has status Success. After the final response the<br />
Veradius AE releases the association.<br />
The contents of the received worklist are compared with the contents of the previous<br />
worklist. In case there are any changes, the Veradius patient file is updated. A unique<br />
match of the following attributes identifies a worklist entry.<br />
© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />
XPS031-090132.02 Veradius R1.1 6 Jul 2009