03.08.2013 Views

Equipment Layout - Philips

Equipment Layout - Philips

Equipment Layout - Philips

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Allura Xper Biplane FD20/10, 20/20 w/Pivot<br />

Preferred Room <strong>Layout</strong><br />

The layout shown below is based upon a typical equipment configuration with full movement range and should be considered as a general<br />

design guideline. Site conditions, application requirements, customer preferences, and/or equipment configuration may significantly impact<br />

suite design and equipment layout. It is recommended to request site-specific drawings from a <strong>Philips</strong> representative early in the design<br />

process.<br />

16'-3"<br />

[4954mm]<br />

Centerline<br />

Clea / Table<br />

6'-0"<br />

[1828mm]<br />

7'-9"<br />

[2362mm]<br />

ATY<br />

11'-4"<br />

[3454mm]<br />

SP<br />

<strong>Equipment</strong> closet doors<br />

to be hung from ceiling<br />

mounted track<br />

PB1<br />

17'-3"<br />

[5258mm]<br />

0 1' 2' 4'<br />

8'<br />

12'<br />

0<br />

.5m<br />

25'-3"<br />

[7695mm]<br />

Clea Floorplate<br />

Reference Line 4'-11<br />

AD7 Table Floorplate<br />

Centerline<br />

Recommended<br />

Minimum<br />

2'-8"<br />

3 8 "<br />

3'-0"<br />

[914mm]<br />

[1507mm]<br />

1m<br />

2m<br />

MSA<br />

TV<br />

MAV<br />

<strong>Equipment</strong> <strong>Layout</strong><br />

Required Ceiling Height: 9' - 9 5<br />

16", + 3<br />

16" / -0 (2980mm, +4mm / -0)<br />

Ceiling Height measured from top of Clea floorplate to bottom of Unistrut.<br />

4m<br />

7'-0"<br />

[2134mm]<br />

IH<br />

16'-3 1 8 "<br />

[4954mm]


Furnished and installed by <strong>Philips</strong><br />

B Furnished by customer/contractor and installed by customer/contractor<br />

C Installed by customer/contractor<br />

D Furnished by <strong>Philips</strong> and installed by contractor<br />

E Existing<br />

F Future<br />

G Optional item furnished by <strong>Philips</strong><br />

A<br />

<strong>Equipment</strong> Designation<br />

Description<br />

<strong>Equipment</strong> Legend<br />

Weight<br />

lbs [kg]<br />

Heat Load<br />

Btu/hr [W]<br />

A SP Floor Clea Stand 2513 [1140] 853 [387]<br />

A MSA Angio Diagnost 7 1693 [768] - [-]<br />

A PB1 Larc N Neuro * 1763 [800] 854 [250]<br />

A MG Velara Generator 40E Cabinet 510 [231] 2971 [870]<br />

A 2MG Velara Generator 40E Cabinet 510 [231] 2971 [870]<br />

A MP Peripheral 40E Cabinet 510 [231] 2049 [600]<br />

A 2MP Peripheral 40E Cabinet 510 [231] 2049 [600]<br />

A PBK PDU 4000/UPS - Required in USA 860 [390] 2450 [717]<br />

A MA Mains 40E Cabinet 710 [322] 4439 [1300]<br />

A CY Viewing / Control 126 [58] 567 [166]<br />

A DB Documentation Box - Mounted on Wheels<br />

(Final location to be coordinated with customer<br />

and/or local <strong>Philips</strong> Service.)<br />

176 [80] - [-]<br />

A ATY Exam Room Auxiliary Box 7 [3] 1.7 [.50]<br />

A TV Six LCD Monitor Suspension 636 [288] 1020 [299]<br />

G MAV Mavig Ceiling Track w/ Radiation Shield 167 [76] 350 [102]<br />

G IH Interventional Hardware 73 [33] 2424 [710]<br />

* Larc CN Cardiac also available (FD20/10 only)


Environmental Requirements for General <strong>Equipment</strong> Locations<br />

Heating, ventilating, air conditioning requirement for general equipment locations must maintain temperature at 72° +/-5° F<br />

(22° +/- 3° C) and non-condensing relative humidity at 20% - 80%.<br />

<strong>Equipment</strong>'s designed airflow is from bottom to top and front to back. Please design the air handling in the rack cabinet<br />

equipment area accordingly.<br />

Power Requirements<br />

Velara Generator with PDU 4000<br />

Power Output: 100 kW<br />

Supply Configuration: 3 phase, 3 wire power and ground. Delta or Wye.<br />

3 phase, 4 wire power with neutral + ground, Wye.<br />

Nominal Line Voltage: 380, 400, 440, 460, or 480 VAC, 60 Hz<br />

Branch Power Requirement: 225 kVA<br />

Circuit Breaker: 3 pole, 125 A<br />

Remote Service Diagnostics<br />

Medical Imaging equipment to be installed by <strong>Philips</strong> is equipped with a service diagnostic feature which allows for remote and<br />

on-site service diagnostics. To establish this feature, a RJ45 type Ethernet 10/100/1000 Mbit network connector must be<br />

installed. Access to customer's network via their remote access server is needed for Remote Service Network (RSN)<br />

connectivity. All costs with this feature are the responsibility of the customer.<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009.<br />

All rights reserved.<br />

Reproduction in whole or in part is<br />

prohibited without prior written consent of<br />

the copyright holder.<br />

Rev. 09.00

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!