<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 52 of 78<br />

4.4.2. Parameters<br />

The specification of important operational parameters, their default value and range ( if<br />

configurable) is specified here.<br />

Table 53: Configuration Parameters Table<br />

Parameter Configurable Default Value<br />

General Parameter<br />

Maximum PDU receive size No -<br />

Maximum PDU send size Yes 16384<br />

Maximum number of simultaneous associations Yes 2<br />

Artim TimeoutSpecifies the time in seconds of the ARTIM (Association<br />

Request/Reject/Release TIMer).<br />

Allowed values:<br />

0: unlimited waiting time0 < n: real time in seconds<br />

Yes 60 [seconds]<br />

Automatic Association Timeout<br />

Specifies the association inactivity timeout in seconds after<br />

which the association is closed automatically.<br />

Allowed values:<br />

-1: immediate timeout0: unlimited waiting time0 < n: real time in<br />

seconds<br />

Yes 0 [unlimited]<br />

Transfer Syntax support: ILE, ELE, EBE<br />

Storage Specific Parameters<br />

Yes ILE, ELE, EBE<br />

Automatic export to a configurable destination Yes -<br />

Background Air Suppression on/off Yes on<br />

Storage Commitment Specific Parameters<br />

Storage Commit Max Reply Waiting Time<br />

Specifies the time in seconds that is waited for a storage<br />

commitment event report message. After this time the association will<br />

be terminated<br />

Allowed values:<br />

-1: immediate timeout<br />

0: unlimited waiting time0 < n: real time in seconds<br />

Yes -1<br />

[asynchronous]<br />

RIS query timeout<br />

Basic Worklist Management Specific Parameters<br />

Yes 240 [minutes]<br />

Specifies the time after which the query is automatically<br />

abortedAllowed values: 1- 300 minutes<br />

Background broad query time interval<br />

Yes 0 [no broad<br />

Specifies the time until the background query will be repeated.<br />

Allowed values:<br />

0: no broad query0 < n: real time in minutes<br />

query]<br />

Print Management Specific Parameters<br />

Automatic print to a configurable destination Yes -<br />

© Koninklijke <strong>Philips</strong> Electronics N.V.<br />

Document Number: XPR 031-090167.00 MammoDiagnost DR 1.0 18-March-2009