DICOM Conformance Statement Page 34 of 78 4.2.1.3.4.1. Description and Sequencing of Activities Figure 6: (Real World) Activity - Image Export Export means that MammoDiagnost DR 1.0 stores images without Storage Commitment. This RWA may be initiated in two ways: • Manually in the viewer, after clicking the Store button the ELEVA AE will store the selected images at the selected Storage SCP. • Automatically during an examination, after clicking the Confirm button the ELEVA AE will automatically store the related images of the performed procedure step at the configured Storage SCP. The ELEVA AE will request an association with the remote Storage SCP for the applicable Storage SOP classes. After accepting the association the ELEVA AE will send the store request, wait for response, and then release the association. The store response status may be inspected on the UI. Depending on the status of the store the ELEVA AE may queue store requests for retries. The queued store requests can be cancelled from the UI. 4.2.1.3.4.2. Proposed Presentation Contexts The presentation contexts are defined in the next table. Table 31: Proposed Presentation Contexts for (Real-World) Activity – Image Export Presentation Context Table Abstract Syntax Transfer Syntax Name UID Name List UID List Computed Radiography Image Storage SOP Class Digital Mammography X- Ray Image Storage - 1.2.840.10008.5.1.4.1.1.1 1.2.840.10008.5.1.4.1.1.1.2 Explicit VR Big Endian 1.2.840.10008.1.2.2 Explicit VR Little Endian 1.2.840.10008.1.2.1 Implicit VR Little Endian 1.2.840.10008.1.2 Explicit VR Big Endian 1.2.840.10008.1.2.2 Explicit VR Little Endian 1.2.840.10008.1.2.1 Exten ded Role Negoti ation SCU None SCU None © Koninklijke Philips Electronics N.V. Document Number: XPR 031-090167.00 MammoDiagnost DR 1.0 18-March-2009
DICOM Conformance Statement Page 35 of 78 Presentation Context Table Abstract Syntax Transfer Syntax Exten Name UID Name List UID List Pres. SOP Implicit VR Little Endian 1.2.840.10008.1.2 Digital Mammography X- 1.2.840.10008.5.1.4.1.1.1.2.1 Explicit VR Big Endian 1.2.840.10008.1.2.2 Ray Image Storage - Proc. SOP Explicit VR Little Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2 Role ded Negoti ation SCU None Note: Only the MG export provides the full set of mammography data attributes (e.g. the IHE MAMMO set), and CR is merely a fall back option. By default, all images are DICOM Stored according to the SOP Class Digital Mammography X-Ray Image storage - for Presentation attributes that are undefined for MG Images are stored in private attributes. The Digital Mammography X-Ray Image IOD is used in two SOP Classes as defined in the DICOM Standard, a SOP Class for storage of images intended for Presentation and a SOP Class for storage of images intended for further Processing before presentation. These are distinguished by their SOP Class UID and by the Enumerated Value of the mandatory Attribute in the DX Series Module, Presentation Intent Type (0008,0068). The SOP Class Digital Mammography X-Ray Image storage - for Processing will be exported if the hospital wants that the for Processing DICOM information is also sent to the archive and/or a CAD device is connected for scanning the "for Processing" images for calcification/nodes. The Numbering Scheme shall support 'Hanging Protocols' of PACS systems & Viewing Stations: 1. The Series Number shall start with 1 for the first Series of every Study Instance, identified by StudyInstanceUID. 2. The Series Number shall increase by 1 for every new Series Instance within the same Study Instance, by the timely order, the Series Instances are created. 3. The Image Number shall start with 1 for every new Series Instance. 4. The Image Number shall increase by 1 for every new Image Instance within the same SeriesInstance, by the timely order, the Images are created. For Digital Mammography X-Ray image Storage SOP Classes is in the DICOM Standard defined: • The Digital Mammography X-Ray Image IOD specifies an image that has been created by a digital mammography imaging device. • A MG image shall consist of the result of a single X-Ray exposure, in order to ensure that the anatomical and orientation attributes are meaningful for the image, permitting safe annotation, appropriate image processing and appropriate dissemination. 4.2.1.3.4.3. SOP Specific Conformance for Storage SOP Classes This section includes the SOP specific behavior, i.e. error codes, error and exception handling, time-outs, etc. Behavior of an Application Entity SOP class is summarized as shown in next Table. The standard as well as the manufacturer specific status codes and their corresponding behavior are specified. © Koninklijke Philips Electronics N.V. Document Number: XPR 031-090167.00 MammoDiagnost DR 1.0 18-March-2009
- Page 1 and 2: DICOM Conformance Statement © Koni
- Page 3 and 4: DICOM Conformance Statement Page 3
- Page 5 and 6: 2. TABLE OF CONTENTS DICOM Conforma
- Page 7 and 8: 3. INTRODUCTION DICOM Conformance S
- Page 9 and 10: DICOM Conformance Statement Page 9
- Page 11 and 12: DICOM Conformance Statement Page 11
- Page 13 and 14: DICOM Conformance Statement Page 13
- Page 15 and 16: DICOM Conformance Statement Page 15
- Page 17 and 18: DICOM Conformance Statement Page 17
- Page 19 and 20: DICOM Conformance Statement Page 19
- Page 21 and 22: DICOM Conformance Statement Page 21
- Page 23 and 24: DICOM Conformance Statement Page 23
- Page 25 and 26: DICOM Conformance Statement Page 25
- Page 27 and 28: Service Status DICOM Conformance St
- Page 29 and 30: DICOM Conformance Statement Page 29
- Page 31 and 32: DICOM Conformance Statement Page 31
- Page 33: DICOM Conformance Statement Page 33
- Page 37 and 38: DICOM Conformance Statement Page 37
- Page 39 and 40: DICOM Conformance Statement Page 39
- Page 41 and 42: DICOM Conformance Statement Page 41
- Page 43 and 44: DICOM Conformance Statement Page 43
- Page 45 and 46: DICOM Conformance Statement Page 45
- Page 47 and 48: DICOM Conformance Statement Page 47
- Page 49 and 50: DICOM Conformance Statement Page 49
- Page 51 and 52: 4.3. Network Interfaces DICOM Confo
- Page 53 and 54: 5. MEDIA INTERCHANGE DICOM Conforma
- Page 55 and 56: DICOM Conformance Statement Page 55
- Page 57 and 58: 7. SECURITY 7.1. Security Profiles
- Page 59 and 60: 8. ANNEXES OF APPLICATION "ELEVA" 8
- Page 61 and 62: DICOM Conformance Statement Page 61
- Page 63 and 64: DICOM Conformance Statement Page 63
- Page 65 and 66: DICOM Conformance Statement Page 65
- Page 67 and 68: DICOM Conformance Statement Page 67
- Page 69 and 70: DICOM Conformance Statement Page 69
- Page 71 and 72: DICOM Conformance Statement Page 71
- Page 73 and 74: DICOM Conformance Statement Page 73
- Page 75 and 76: View Position 0018,5101 CS VNAP DIC
- Page 77 and 78: Number of Films 2100,0170 IS DICOM
<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 35 of 78<br />
Presentation Context Table<br />
Abstract Syntax Transfer Syntax Exten<br />
Name UID Name List UID List<br />
Pres. SOP Implicit VR Little Endian 1.2.840.10008.1.2<br />
Digital Mammography X- 1.2.840.10008.5.1.4.1.1.1.2.1 Explicit VR Big Endian 1.2.840.10008.1.2.2<br />
Ray Image Storage -<br />
Proc. SOP<br />
Explicit VR Little Endian<br />
Implicit VR Little Endian<br />
1.2.840.10008.1.2.1<br />
1.2.840.10008.1.2<br />
Role<br />
ded<br />
Negoti<br />
ation<br />
SCU None<br />
Note: Only the MG export provides the full set of mammography data attributes (e.g.<br />
the IHE MAMMO set), and CR is merely a fall back option.<br />
By default, all images are <strong>DICOM</strong> Stored according to the SOP Class Digital<br />
Mammography X-Ray Image storage - for Presentation attributes that are undefined<br />
for MG Images are stored in private attributes.<br />
The Digital Mammography X-Ray Image IOD is used in two SOP Classes as defined<br />
in the <strong>DICOM</strong> Standard, a SOP Class for storage of images intended for Presentation<br />
and a SOP Class for storage of images intended for further Processing before<br />
presentation.<br />
These are distinguished by their SOP Class UID and by the Enumerated Value of the<br />
mandatory Attribute in the DX Series Module, Presentation Intent Type (0008,0068).<br />
The SOP Class Digital Mammography X-Ray Image storage - for Processing will be<br />
exported if the hospital wants that the for Processing <strong>DICOM</strong> information is also sent<br />
to the archive and/or a CAD device is connected for scanning the "for Processing"<br />
images for calcification/nodes.<br />
The Numbering Scheme shall support 'Hanging Protocols' of PACS systems &<br />
Viewing Stations:<br />
1. The Series Number shall start with 1 for the first Series of every Study Instance,<br />
identified by StudyInstanceUID.<br />
2. The Series Number shall increase by 1 for every new Series Instance within the<br />
same Study Instance, by the timely order, the Series Instances are created.<br />
3. The Image Number shall start with 1 for every new Series Instance.<br />
4. The Image Number shall increase by 1 for every new Image Instance within the<br />
same SeriesInstance, by the timely order, the Images are created.<br />
For Digital Mammography X-Ray image Storage SOP Classes is in the <strong>DICOM</strong><br />
Standard defined:<br />
• The Digital Mammography X-Ray Image IOD specifies an image that has been<br />
created by a digital mammography imaging device.<br />
• A MG image shall consist of the result of a single X-Ray exposure, in order to<br />
ensure that the anatomical and orientation attributes are meaningful for the<br />
image, permitting safe annotation, appropriate image processing and<br />
appropriate dissemination.<br />
4.2.1.3.4.3. SOP Specific <strong>Conformance</strong> for Storage SOP Classes<br />
This section includes the SOP specific behavior, i.e. error codes, error and exception<br />
handling, time-outs, etc.<br />
Behavior of an Application Entity SOP class is summarized as shown in next Table.<br />
The standard as well as the manufacturer specific status codes and their<br />
corresponding behavior are specified.<br />
© Koninklijke <strong>Philips</strong> Electronics N.V.<br />
Document Number: XPR 031-090167.00 MammoDiagnost DR 1.0 18-March-2009