3.1 2D QC-3V phantom Projections<br />

Figure 2: Contrast resolution phantom<br />

3. Results<br />

Figures 3a and 3b illustrate the QC-3V phantom images acquired with receptors 1 and 2. The QC<br />

images were analyzed by the PIPS-pro 3 software to compute the CNR, f50, f40, f30, and are shown in<br />

Table-1.<br />

Receptor Dose (cGy) CNR f50 (lp/mm) f40 (lp/mm) f30 (lp/mm)<br />

1 0.025 43 0.39 0.50 0.64<br />

2 0.035 69 0.41 0.53 0.66<br />

Table-1: Computed spatial resolution and CNR