23.07.2013 Views

Soft Tissue Visualization Using a Highly Efficient Megavoltage Cone ...

Soft Tissue Visualization Using a Highly Efficient Megavoltage Cone ...

Soft Tissue Visualization Using a Highly Efficient Megavoltage Cone ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Soft</strong> <strong>Tissue</strong> <strong>Visualization</strong> <strong>Using</strong> a <strong>Highly</strong> <strong>Efficient</strong> <strong>Megavoltage</strong><br />

<strong>Cone</strong> Beam CT Imaging System<br />

Farhad A. Ghelmansarai, Ali Bani-Hashemi, Jean Pouliot*, Ed Calderon, Paco<br />

Hernandez, Matthias Mitschke, Michelle Aubin*, Kara Bucci*<br />

Siemens Medical Solutions Inc., Concord, CA, USA<br />

*Univ. of California at San Francisco, San Francisco, CA<br />

ABSTRACT<br />

Recent developments in two-dimensional x-ray detector technology have made volumetric <strong>Cone</strong> Beam CT<br />

(CBCT) a feasible approach for integration with conventional medical linear accelerators. The requirements<br />

of a robust image guidance system for radiation therapy include the challenging combination of soft tissue<br />

sensitivity with clinically reasonable doses. The low contrast objects may not be perceptible with MV<br />

energies due to the relatively poor signal to noise ratio (SNR) performance.<br />

We have developed an imaging system that is optimized for MV and can acquire <strong>Megavoltage</strong> CBCT<br />

images containing soft tissue contrast using a 6MV x-ray beam. This system is capable of resolving relative<br />

electron density as low as 1% with clinically reasonable doses ranging from 3 to 16 cGy. There are many<br />

factors such as image noise, x-ray scatter, improper calibration and acquisitions that have a profound effect<br />

on the imaging performance of CBCT and in this study attempts were made to optimize these factors in<br />

order to maximize the SNR. A QC-3V phantom was used to determine the contrast to noise ratio (CNR)<br />

and f50 of a single 2-D projection. The computed f50 was 0.43 lp/mm and the CNR for a radiation dose of<br />

0.02cGy was 43. Clinical <strong>Megavoltage</strong> CBCT images acquired with this system demonstrate that<br />

anatomical structures such as the prostate in a relatively large size patient are visible using clinically<br />

acceptable radiation doses.<br />

Key Words: <strong>Megavoltage</strong> cone beam CT, Electronic portal imaging device, <strong>Megavoltage</strong> soft tissue<br />

imaging.<br />

1. Introduction<br />

The goal of the radiotherapy is to deliver the highest possible dose to the tumor while minimizing the dose<br />

to the surrounding tissue. Conformal Intensity modulated radiotherapy (IMRT) and conformal radiotherapy<br />

have been developed to achieve these goals. However, the benefits of these techniques are limited when<br />

large margins have to be used to accommodate uncertainties in target and normal tissue at the time of<br />

treatment. One possible solution is the computed tomography (CT) integrated with the treatment machine.<br />

There are a few approaches available to CT imaging in the treatment position. One such approach<br />

integrates a KV x-ray tube and an a-Si flat panel detector that is mounted on the C-arm gantry of a linear<br />

accelerator. Second approach is to use of the treatment MV beam to obtain CT images either in fan beam or<br />

cone beam scanning geometry.<br />

The use of MV photons to obtain CT images results in a greater absolute dose deposition per photon<br />

compared to KV photons. Therefore, in order to maintain comparable doses for MVCT and KVCT, the<br />

number of MV photons incident on the patient must be considerably reduced and this reduction decreases<br />

the signal to noise ratio. One mitigating factor for MVCT is that the MV photons are more penetrating than<br />

KV photons. Thus, a greater fraction of incident MV photons will travel through the patient and reach the<br />

detector. This effect will provide some compensation for the reduced number of incident photons,<br />

especially for larger patients. Unfortunately, this effect is countered by the increased penetration of the<br />

photons through the detector.


The ability of an x-ray imaging system to differentiate soft tissues is affected by the difference in<br />

attenuation coefficient at a given energy between the contrasting objects. For KV CBCT imaging the<br />

contrast changes with energy, because both the photoelectric and Compton effects contributes significantly<br />

to attenuation. For MV energies the Compton effect is by far the dominant interaction process, especially<br />

for material with low atomic number (Z), in which the attenuation coefficient is proportional to electron<br />

density and is nearly independent of Z. Object contrast is therefore constant over a wide range of energies<br />

in the MV regime. Since the attenuation coefficients of KV photons are higher than MV photons in soft<br />

tissue, the percent density contrasts will be easier to detect at KV energies. Thus, while the higher<br />

attenuation of KV photons results in greater transmission losses, it also benefits KV contrast by the squared<br />

ratio of the mass attenuation coefficient for KV and MV energies 1 .<br />

In this paper we present an implementation of a highly efficient MV CBCT imaging system that can<br />

resolve soft tissue contrast with clinically acceptable doses.<br />

2. Methods and Materials<br />

2.1 Timing Interface for <strong>Cone</strong> Beam Acquisition Mode<br />

The timing interface uses the Radiation-on (Rad-on) signal of linear accelerator (linac) to generate trigger<br />

pulses that externally control the frame readout of flat panel imager. In this mode, a-Si sensors integrate the<br />

signal during the radiation interval of each projection and the data readout is performed at the end of each<br />

projection. The detector is set in the external mode. Five trigger pulses (5 frames scan) are sent to the<br />

external trigger input of the flat panel detector prior to the start of first projection to force sensor discharge<br />

by reading 5 frames in absence of radiation. This eliminates the dark current accumulation prior to the start<br />

of the first projection. No readout is allowed during the projection interval. Avoiding readout during<br />

exposure interval improves the image SNR by reducing the effect of the electronic readout noise, and also<br />

removes the linac pulsing artifacts from the image.<br />

Rad-on<br />

Trigger<br />

2.2 Image Receptor<br />

5 Refresh Pulses<br />

First<br />

Projection<br />

Projection period<br />

Second<br />

Projection<br />

Figure-1: Timing Interface for <strong>Cone</strong> Beam Mode<br />

Third<br />

Projection<br />

Data Projection 1 Data Projection 2 Data Projection 3


Two image receptors that were used for this study contained two components; a metal build up<br />

plate/phosphor screen and a flat panel imager. The flat panel imager had an active area of 41 x 41 cm 2 , and<br />

contained 1024 x1024 elements of 0.4 mm pitch. The fastest frame rate was 7 frames per second and the<br />

output data was 16-bit. Flat panel detectors generate frames either using “Free Running” or “External<br />

Triggered”. Free running means that the detector sends out continuous frames according to the selected<br />

frame time. External triggered means that there is an external trigger which allows the generation of the<br />

maskable interrupt request.<br />

The phosphor screen of one receptor (receptor 1) was a 133 mg/cm 2 Gd 2O 2S screen (Lanex Fast) and for<br />

the receptor 2 was an 800 µm CsI (TI) scintillator screen. Both receptors were using a 1mm copper buildup<br />

plate. The charge amplifier capacitance of receptors 1 and 2 were 5pF and 1pF, respectively.<br />

2.3 <strong>Cone</strong> Beam Image Acquisitions and Corrections<br />

<strong>Cone</strong> beam CT starts by acquiring a large number of projection images acquired as the X-ray source<br />

follows an orbital path. At every angular increment, i.e., every degree, an image is acquired. 201<br />

projections were acquired over 201 degrees (1° rotation between projections) around the subject. The<br />

gantry speed was set to 109° per minute.<br />

The Siemens PRIMUS® linear accelerator parameters were adjusted to allow the delivery of a small<br />

fraction of a Monitor Unit (MU) per image. A very precise dose control was achieved by loading the<br />

accelerator by very small amount of electrons per pulse. In the current design, the injection is phased with<br />

the RF power at startup or during re-phasing of the injector at the programmed angular increment. With<br />

normal beam parameters, the phasing or re-phasing of the injection generates a shift in the resonant<br />

frequency of the accelerator. Because of the lag of the automatic frequency control (AFC), the beam<br />

formation can be significant (hundreds of milliseconds). In order to improve the beam formation, the beam<br />

parameters were adjusted such that the dose per pulse was reduced to approximately 10% of normal, and<br />

the pulse repetition frequency (PRF) was adjusted to 250 pps. The reduced loading resulted in an<br />

insignificant shift in the resonant frequency of the accelerator, resulting in significant improvement of the<br />

beam formation (less than a few ms). The increased PRF improved the granularity of the control system,<br />

the significance of every dose pulse was reduced, the predictability of the system increased, and the beam<br />

formation time decreased to only several milliseconds.<br />

Three corrections were performed; Offset, Gain, and Defected pixel correction. Offset correction is used to<br />

correct the dark current of each pixel for a specified frame or integration time. The integration time for each<br />

projection is equal to the projection period (figure-1), and therefore, the Offset correction images were<br />

acquired with the frame time that is equal to the projection period.<br />

The Gain correction is used to homogenize the flat field image intensity. The Defected pixel correction<br />

allows a software repair of defected pixels to enhance image quality. Improper pixel values are replaced<br />

with the averaged value of the adjacent pixels, while defected pixels are not averaged.<br />

2.4 Preprocessing by Resolution/Dose Adaptive Filter<br />

A unique adaptive noise removal filter was used for image enhancement 2 . The filter utilizes pixel-wise<br />

nonlinear and adaptive techniques to estimate the additive noise power before filtering. The characteristics<br />

of the filter are adapted to the input X-ray flux and image spatial resolution. Filtering is minimal to the<br />

images that receives high X-ray flux, and increases with the decreasing flux. To minimize the loss of<br />

resolution, minimum filtering is applied to the area of an image with large variance to preserve the edges<br />

and other high frequency components of an image 2 .<br />

Before data normalization and reconstruction, the corrected projection images (Offset, Gain, and Defected<br />

pixel) were processed by the resolution/dose adaptive filter using a local matrix size of 8x8.<br />

2.5 Data Normalization and Reconstruction


The corrected (Offset, Gain, Dead pixel) and pre-processed (resolution/dose adaptive filtering) projection<br />

images were normalized using the pixel data in the open area of the image.<br />

///////<br />

Volumetric reconstruction is accomplished with using an extension of Feldkamp filtered back projection in<br />

cone beam geometry 5 . The reconstruction time for generating 256 3 CT data was ??<br />

2.6 Phantoms for Image Quality Analysis<br />

The followings are the list of phantoms that were used and all were positioned at isocenter. The source to<br />

imager distance (SID) was 140cm for all phantom images.<br />

• A QC-3V phantom 3 was used to determine the spatial resolution and contrast to noise ratio (CNR)<br />

of 2D projection images.<br />

• A ball phantom with a diameter of 16cm consisting of water-equivalent plastic was used for<br />

contrast/spatial resolution experiments. The 1cm thick slice in the middle of the ball contains<br />

several tissue equivalent inserts that are 3 cm in diameter (Figure-2). There are also some inserts<br />

with smaller diameters (1, 0.5, and 0.25 cm). There are air filled holes with diameters of 1, 1.5, 2,<br />

and 3mm within the middle slice of phantom for spatial resolution analysis.<br />

• A Quality Assurance System for Advanced Radiotherapy (QUASAR) body phantom 4 that<br />

contains Electron Density (ED) rods of lung inhale, polyethylene, water equivalent, trabecular<br />

bone, and dense bone with EDs of 0.190, 0.945, 1.002, 1.117, and 1.512; respectively. The ED of<br />

acrylic is 1.137.<br />

• An anthropomorphic head phantom with implanted gold seeds (1mm in diameter and 3mm long)


3.1 2D QC-3V phantom Projections<br />

Figure 2: Contrast resolution phantom<br />

3. Results<br />

Figures 3a and 3b illustrate the QC-3V phantom images acquired with receptors 1 and 2. The QC<br />

images were analyzed by the PIPS-pro 3 software to compute the CNR, f50, f40, f30, and are shown in<br />

Table-1.<br />

Receptor Dose (cGy) CNR f50 (lp/mm) f40 (lp/mm) f30 (lp/mm)<br />

1 0.025 43 0.39 0.50 0.64<br />

2 0.035 69 0.41 0.53 0.66<br />

Table-1: Computed spatial resolution and CNR


(a): 0.025cGy receptor 1 (b): 0.035cGy receptor 2<br />

3.2 Reconstructed Images<br />

3.2.1 Phantom Images<br />

Figure 3: QC-3V phantom images<br />

Figure 4 shows the QUASAR and contrast/spatial resolution (ball) phantom images that were acquired<br />

with receptor 1. The total delivered dose was 18cGy and no preprocessing filtering used. From Figure<br />

4(a) 2% electron density variation between acrylic and trabecular bone is resolvable without applying<br />

the adaptive filtering. Applying the resolution/dose adaptive filter will reduce the standard deviation in<br />

pixel intensities by a factor of ~0.5. <strong>Using</strong> the ratio of pixel intensity difference (due to density<br />

difference) to root mean square uncertainty in pixel intensities of receptor 1, it can be estimated that<br />

the receptor 2 due to its higher CNR compared to receptor 1 and in combination with preprocessing<br />

adaptive filtering is capable of resolving 1% contrast with radiation doses as low as 10cGy.<br />

The left image on figure 5 illustrates that muscle insert with contrast of 0.5% is resolvable with<br />

receptor 1 using 163cGy dose and without applying dose/resolution adaptive filtering. By using only<br />

the odd projections used for the previous reconstruction (163cGy) that corresponds to the dose value of<br />

81.5cGy (center image of figure 5), the 0.5% contrast is still detectable. The right hand image of figure<br />

5 shows that the minimum required dose in which the 0.5 % contrast becomes discernible with<br />

receptor 1 and by applying the adaptive filter is ~41MU (threshold of detection of 0.5% contrast). Note<br />

that all figure 5 images have the cupping artefacts that are due to the scattered, and no scatter/beam<br />

profile corrections have been applied to these images.<br />

Figures 4(b) and 6(a) illustrate that the 2mm air holes and gold seeds implanted (1mm x 3mm) within<br />

a head phantom are resolvable.<br />

Figure 6 (a) shows axial, coronal and sagittal images of a head phantom that is acquired with receptor<br />

2 by delivering only 1cGy. Figure 6 (b) displays the raw image (no adaptive filtering) of the 2D<br />

projection of this phantom acquired with 0.005cGy. The mean/standard deviation (SD) ratios of this<br />

2D projection (on open field regions) before and after applying dose/resolution adaptive filter are 26<br />

and 68, respectively. This indicates the possibility of the use of receptor 2 for 4D-CT.


(a) 18cGy QUASAR (b) 18cGy contrast/spatial resolution ball phantom<br />

Figure 4: QUASAR and contrast/spatial resolution phantom images with receptor 1 (without preprocessing<br />

adaptive filtering)<br />

Figure 5: Contrast/spatial resolution phantom images with receptor 1. Left: 163cGy, Center: 81.5cGy,<br />

Right: 41cGy with adaptive filtering<br />

(a) 1cGy reconstructed images (b) Projection image with 0.005cGy


3.2.2 Patient Images<br />

Figure 6: Receptor 1Head phantom images with gold seeds<br />

All patient images were acquired at SID of 145cm. Figure 7 shows a pelvis image acquired with<br />

receptor 1. Total dose was 16cGy and preprocessing adaptive filter was applied to the projection<br />

images. The contribution of beam profile and scatter effects that cause shading artefacts are minimized<br />

by obtaining empirical data (from different anatomies, SID, and patient sizes) and linking to the<br />

correction images. Figure 8 compares the axial view of 8cGy MV CBCT with the planning CT image<br />

and shows that the most of the soft tissue structures such as prostate are detectable with MV CBCT.<br />

Figures 9 and 10 depict axial, coronal and sagittal head/neck images that were acquired with receptors<br />

1 and 2 by delivering 16 and 6cGy, respectively. Optic nerves are detectable in both figures. Figure 11<br />

compares the lateral views of MV CBCT images acquired for three different patients over the past year<br />

and shows the improvement in image quality.<br />

One distinct advantage of MV CBCT imaging is that image quality is maintained even when high-Z<br />

materials such as tooth fillings, dental implants, fiducial markers or hip replacements are present. For<br />

kV imaging, these objects create strong artifacts. This is illustrated in the figure 12.<br />

Figure 7: Pelvis MV CBCT with receptor 1 using 16cGy


Figure 8: Axial view of<br />

the pelvic area.<br />

Left: regular CT.<br />

Center; MV CBCT<br />

using 8 cGy. Right: an<br />

insert displays the<br />

MVCBCT image superimposed<br />

on CT<br />

Figure 9: Head/neck MVCB CT with receptor 1 using 16cGy


Figure 10: Head/neck MVCB CT with receptor 2 using 6cGy<br />

Figure 11: The lateral views of MV CBCT images acquired<br />

for three different patients over the past year, which show the<br />

improvement in image quality and required dose.<br />

Left: first MV CBCT clinical image of a Head&Neck patient<br />

Panel RID 1640, dose=8 cGy).<br />

Center: receptor 1, Dose=8 cGy.<br />

Right: receptor 2, Dose=3 cGy.<br />

Figure 12: Images of a Head & Neck patient with dental implants. Left: an<br />

axial view of the planning kV CT shows large artifacts due to the presence<br />

of the metallic objects. Right: the same anatomical area is shown on an<br />

axial view of a MV CBCT acquired with receptor 2. The dental implants<br />

are visible as well as other surrounding normal tissues which are not<br />

visible on the kV CT.


Figure 13 shows neck/chest MVCT axial, coronal and sagittal images that were acquired with<br />

receptors 1 and by delivering 13 cGy.<br />

Figure 13: Neck/chest MV CBCT with receptor 1 using 13cGy<br />

4. Summary<br />

We have developed an imaging system that is optimized for MV and can acquire <strong>Megavoltage</strong> CBCT<br />

images containing soft tissue contrast using a 6MV x-ray beam with irradiation in range of 3 to 16cGy.<br />

The main components of this system are a high quantum efficient flat panel detector, readout interface<br />

with linac, calibration/corrections process, image preprocessing with dose/resolution adaptive filter,<br />

and scatter corrections. It can resolve a contrast resolution of 1%, gold seeds of 1mm x 3mm (1mm in<br />

diameter and 3mm in length), and air structures of 2mm in diameter.<br />

Clinical MV CBCT images revealed that structures such as prostate and optical nerve are visible with<br />

irradiation of 3 to 8cGy. This system (using receptor 1) can also be used for routine portal imaging<br />

applications without risk of saturation for high dose/high energy treatment/verification imaging, or<br />

dosimetry applications. This system can also deliver good quality MV CBCT images with 1 cGy that<br />

opens the possibility of the use of this system for 4D-CT.<br />

REFERENCES<br />

1. H. Barrett, S. Gordon, and R. Hershel, Statistical limitations in transaxial tomography, Comput. Biol.<br />

Med. 6, 307-323, 1976.<br />

2. F. Ghelmansarai, P. Virador, M. Mitschke, A new dose and resolution adaptive filter, AAPM<br />

(American Association of Physicist in Medicine) 46 th Annual Meeting, 2004.<br />

3. R. Rajapakshe, K. Luchka, and S. Shalev, A quality control test for electronic portal imaging devices,<br />

Med. Phys. 23 (7) 1237-1244, 1996.


4. Modus Medical Devices Inc., QUASAR Body Phantom User’s Guide, May 2001.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!