Location Requirements<br />

Chapter 2: Prerequisites<br />

Note: The Biosafety Cabinet should be located away from traffic patterns,<br />

doors, fans, ventilation registers, fume hoods and any other air-handling<br />

devices that could disrupt its airflow patterns. All windows in the room<br />

should remain closed. Figure 2-1 shows the preferred location for the<br />

Biosafety Cabinet.<br />

Preferred location<br />

Air register blocked or redirected<br />

to prevent cabinet disruption<br />

Figure 2-1<br />

Alternate location<br />

Product Service 1-800-522-7658 3