Chapter 6: Maintaining The Cabinet<br />
• Check the UV and fluorescent light hourmeters, and record their<br />
readings in an operational log.<br />
• Check the prefilter for cleanliness; replace if needed.<br />
• All weekly activities.<br />
Semiannually or Annually<br />
• Have the cabinet re-certified by a qualified certification technician.<br />
• All monthly activities.<br />
• Replace the prefilter.<br />
Service Operations<br />
Work Surface Removal:<br />
Note: The work surface of the cabinet must be thoroughly decontaminated<br />
before removing it.<br />
1. Lift the front edge of the work surface straight up by grasping the knob<br />
handles at either front corner.<br />
2. Pull the work surface straight out, letting its rear edge rest on the center<br />
support underneath.<br />
3. Reinstall the work surface by resting the bottom on the center rail while<br />
pushing it back into the cabinet. Be sure to engage the tabs on the back<br />
corners of the work surface with the slots on the rear wall of the work<br />
area.<br />
Front Grille Removal:<br />
Note: The grille must be thoroughly decontaminated before removing it.<br />
1. Remove the work surface as described earlier.<br />
2. At one end of the grille, grip the front of grille with one hand, and the<br />
back with the other hand. Pivot that end of the grille upward and<br />
inward, paralleling the angle of the sash, as shown in Figure 6-1.<br />
3. Pull the other end of the sash up and away from the bottom edge of the<br />
cabinet.<br />
4. Reinstall the grille by reversing the above sequence, ensuring that the<br />
grille properly engages the bottom edge of the cabinet.<br />
Product Service 1-800-522-7658 43