If An Exhaust Error Activates<br />

The most common causes of an Exhaust Error are:<br />

• Exhaust blower is turned off or has lost power.<br />

• Excessive loading of the Exhaust HEPA filter.<br />

• Leakage in the exhaust system.<br />

• Sudden change in room pressure.<br />

Chapter 5: Using The Cabinet<br />

When an Exhaust Error is activated, the Logic can be reset by pressing the<br />

“Blower” button. The unit will restart and run for 15 seconds, and then will<br />

recheck the airflow. If the airflow is still insufficient, the unit will again<br />

show and Exhaust Error alarm.<br />

NOTE: If the Exhaust Error sounds during use, take appropriate action to<br />

prevent contamination of you and others in the area.<br />

NOTE: If the cabinet keeps reporting Exhaust Error, do not use the cabinet<br />

until the problem has been corrected.<br />

If An Airflow Alert Activates<br />

The most common cause of an Airflow Alert is:<br />

• Blockage of the prefilter.<br />

Resetting the Airflow Alert System<br />

The Airflow Alert automatically resets to normal operation once the motor<br />

speed has stabilized.<br />

Working In the Biosafety Cabinet<br />

Planning<br />

• Thoroughly understand procedures and equipment required before<br />

beginning work.<br />

• Arrange for minimal disruptions, such as room traffic or entry into the<br />

room while the cabinet is in use.<br />

Product Service 1-800-522-7658 39