Chapter 5: Using The Cabinet<br />
UV light<br />
In this mode the UV (germicidal) light is operating. When the<br />
sash is fully closed, the UV light can be programmed to<br />
automatically turn on. When the sash is raised, the light<br />
automatically turns off, to protect the laboratory from UV<br />
irradiation.<br />
Timer<br />
When either the interval or stopwatch timer mode is active, this<br />
icon appears.<br />
Security Lock<br />
When the security lock is activated the keypad does not respond to<br />
any buttons until the proper sequence of keystrokes is entered.<br />
Mute<br />
When this icon is on, the audible alarm is muted for<br />
approximately 5 minutes.<br />
Operating the Sliding Sash<br />
The counterbalanced, anti-racking sash mechanism requires only a few<br />
pounds of force to move the sash up or down. You can open or close the sash<br />
smoothly with one or two hands positioned on either handle.<br />
The sash position alarm and safety interlock system senses the sash position<br />
and acts appropriately. The Biosafety Cabinet has been programmed to<br />
operate at an 8-inch (203 mm) sash opening. Raising the sash above its<br />
operating height will activate the audible and visual alarms. The audible<br />
alarm can be temporarily muted (for approximately five minutes) by<br />
depressing and releasing the Mute/OK button. Closing the sash back to its<br />
operating position will reset the alarm and defeat the muting of the alarm.<br />
The safety interlock system senses when the sash is closed and allows the<br />
optional ultraviolet (UV) lamp to operate only when the sash is closed, to<br />
protect the operator from irradiation.<br />
Product Service 1-800-522-7658 25