Chapter 4: Performance Features & Safety Precautions<br />
Airflow Sensor<br />
An airflow sensor, located above the Exhaust HEPA filter, constantly<br />
monitors the flow of exhaust air out of the cabinet. If the exhaust falls below<br />
a safe level, the control board turns off the cabinet blower, and sounds an<br />
audible and visual alarm. This prevents the escape of hazardous material<br />
from the front of the cabinet, in the event of an exhaust system failure. The<br />
airflow sensor reading is shown on the display as Inflow in feet per minute<br />
(or m/s).<br />
Cabinet Air Intakes (Grilles), Ductwork and Air<br />
Balance Controls<br />
The location, size, and pattern of the grilles at the front and rear of the work<br />
area affect cabinet containment and performance.<br />
Note: Never block or obstruct the grilles of the Biosafety Cabinet.<br />
Ultraviolet (UV) Lamp<br />
The optional UV lamp generates a primary wavelength of light of 254nm. A<br />
secondary emission is in the visible (blue) wavelength, resulting in the<br />
characteristic blue color while operating. UV light with a wavelength of 254<br />
nm is biocidal, primarily by creating thymine dimers in DNA. These dimers<br />
prevent the correct transcription of the DNA into RNA, resulting in cellular<br />
death or viral inactivation. In order to be effective, the UV light must directly<br />
strike the nucleic acid, and its effectiveness can be diminished or negated by<br />
dissolved proteins or metals, or by other UV-opaque substances protecting<br />
the target nucleic acid.<br />
Because of its limitations, UV light should be used as an adjunct to good<br />
surface disinfection practices. In order to get optimum performance from the<br />
UV light, it should be replaced after 6,000 hours of operation or less, and the<br />
exterior surface of the lamp should be kept clean and free of dust.<br />
Note: The Logic Cabinet records the number of hours of operation of the UV<br />
light. You can program in the number of hours (in 100-hour increments) it<br />
will operate before a replacement message is displayed.<br />
Note: UV irradiation is absorbed by the tempered safety glass of the sash.<br />
Independent research has shown that the level of UV irradiation on the<br />
outside of the cabinet’s sash is equal to background radiation levels.<br />
Note: The UV sensitivity of a target organism varies, depending on the UV<br />
output of the lamp, the genus and species of the organism, the medium the<br />
agent is suspended in, etc. Contact the facilities Health and Safety Officer at<br />
your facility for UV light use and recommendations.<br />