NOTE: The drain valve assembly attaches to<br />

the underside of the cabinet bottom.<br />

Apply a light coat of silicone sealant to this<br />

surface of the connector, aligning the three<br />

holes in the connector with the three holes in<br />

the Biosafety Cabinet liner.<br />

Initial Certification<br />

Chapter 3: Getting Started<br />

Prior to use, a qualified certifier should certify all Biosafety Cabinets. Under<br />

normal operating conditions, the Biosafety Cabinet should be recertified at<br />

least annually and when relocated or serviced. The certifier should perform<br />

the following tests, as recommended in ANSI/NSF International Standard<br />

Number 49 in effect when the cabinet was manufactured:<br />

• Downflow Velocity Profile Test<br />

• Inflow Velocity Test<br />

• Airflow Smoke Patterns<br />

• HEPA Filter Leak Test<br />

• Vibration Test *<br />

• Noise Level Test *<br />

• Lighting Intensity Test *<br />

*These tests are user comfort related tests and may be omitted at the<br />

user’s or certifier’s discretion.<br />

Figure 3-4<br />

If you have any questions regarding certification agencies or need assistance<br />

in locating one, contact Labconco’s Product Service Department at 1-800-<br />

522-7658 or 816-333-8811.<br />

Product Service 1-800-522-7658 13