Exhaust System Connections<br />
Chapter 3: Getting Started<br />
The Purifier Logic Series Total Exhaust Biosafety Cabinets are a “Type B2<br />
cabinet,” meaning they direct their HEPA filtered exhaust air out of the<br />
laboratory.<br />
Note: The Logic Total Exhaust Biosafety Cabinet is designed to be<br />
connected to an appropriate exhaust system. Do not attempt to operate<br />
it when it is not connected to an appropriate exhaust system.<br />
NOTE: THE EXHAUST BLOWER CANNOT RUN AT ITS NOMINAL<br />
FLOW RATE WITH THE SASH COMPLETELY CLOSED. IF YOU<br />
CHOOSE TO CLOSE THE SASH COMPLETELY TO USE THE UV<br />
LIGHT FOR SURFACE DISINFECTION, THEN THE EXHAUST<br />
BLOWER MUST BE SHUT OFF, OR ITS FLOW RATE REDUCED 90%<br />
OR MORE FOR PROPER OPERATION. THE LOGIC CONTROL BOARD<br />
HAS DRY CONTACTS THAT CAN BE CONFIGURED TO SIGNAL THE<br />
EXHAUST BLOWER TO TURN ON WHEN THE LOGIC BLOWER IS<br />
ON. FOR FURTHER INFORMATION, CONTACT LABCONCO’S<br />
PRODUCT SERVICE DEPARTMENT. AS AN ALTERNATIVE, A<br />
REMOTE ELECTRICAL SWITCH FOR THE EXHAUST BLOWER CAN<br />
BE INSTALLED NEAR THE LOGIC.<br />
THE EXHAUST CONNECTION IS A SEALED EXHAUST SYSTEM<br />
FROM THE TOP OF THE CABINET TO THE REMOTE BLOWER. THE<br />
EXHAUST SYSTEMS SHOULD BE DEDICATED TO A SINGLE<br />
CABINET. THE CABINET IS EQUIPPED WITH AN EXHAUST FLOW<br />
ALARM, SUCH THAT AN ALARM SOUNDS, AND THE CABINET<br />
BLOWER SHUTS OFF IN THE EVENT OF INSUFFICIENT EXHAUST<br />
FLOW.<br />
If your research involves the use of toxic compounds or volatile materials,<br />
contact your facility’s safety officer or Labconco to ensure that your Purifier<br />
and its exhaust system are compatible with the materials you will be working<br />
with.<br />
Product Service 1-800-522-7658 11