None of the parameters had a significant effect on the results of<br />

the ruggedness test.<br />

9.2 Precision. Many TCLP precision (reproducibility) studies have been<br />

performed, and have shown that, in general, the precision of the TCLP is<br />

comparable to or exceeds that of the EP toxicity test and that method precision<br />

is adequate. One of the more significant contributions to poor precision appears<br />

to be related to sample homogeneity and inter-laboratory variation (due to the<br />

nature of waste materials).<br />

9.2.1 <strong>Metals</strong> - The results of a multi-laboratory study are shown<br />

in Table 6, and indicate that a single analysis of a waste may not be<br />

adequate for waste characterization and identification requirements.<br />

9.2.2 Semi-Volatile Organic Compounds - The results of two<br />

studies are shown in Tables 7 and 8. Single laboratory precision was<br />

excellent with greater than 90 percent of the results exhibiting an RSD<br />

less than 25 percent. Over 85 percent of all individual compounds in the<br />

multi-laboratory study fell in the RSD range of 20 - 120 percent. Both<br />

studies concluded that the TCLP provides adequate precision. It was also<br />

determined that the high acetate content of the extraction fluid did not<br />

present problems (i.e., column degradation of the gas chromatograph) for<br />

the analytical conditions used.<br />

9.2.3 Volatile Organic Compounds - Eleven laboratories<br />

participated in a collaborative study of the use of the ZHE with two waste<br />

types which were fortified with a mixture of VOCs. The results of the<br />

collaborative study are shown in Table 9. Precision results for VOCs tend<br />

to occur over a considerable range. However, the range and mean RSD<br />

compared very closely to the same collaborative study metals results in<br />

Table 6. Blackburn and Show concluded that at the 95% level of significance:<br />

1) recoveries among laboratories were statistically similar, 2)<br />

recoveries did not vary significantly between the two sample types, and 3)<br />

each laboratory showed the same pattern of recovery for each of the two<br />

samples.<br />

10.0 REFERENCES<br />

1. Blackburn, W.B. and Show, I. "Collaborative Study of the Toxicity<br />

Characteristics Leaching Procedure (TCLP)." Draft Final Report, Contract No. 68-<br />

03-1958, S-Cubed, November 1986.<br />

2. Newcomer, L.R., Blackburn, W.B., Kimmell, T.A. "Performance of the<br />

Toxicity Characteristic Leaching Procedure." Wilson Laboratories, S-Cubed, U.S.<br />

EPA, December 1986.<br />

3. Williams, L.R., Francis, C.W.; Maskarinec, M.P., Taylor D.R., and Rothman,<br />

N. "Single-Laboratory Evaluation of Mobility Procedure for Solid Waste." EMSL,<br />

ORNL, S-Cubed, ENSECO.<br />

CD-ROM 1311- 23 Revision 0<br />

July 1992