Precious Metals Recovery LLC RCRA Permit Application Dry Hills ...
Precious Metals Recovery LLC RCRA Permit Application Dry Hills ... Precious Metals Recovery LLC RCRA Permit Application Dry Hills ...
8.2.1 Matrix spikes are to be added after filtration of the TCLP extract and before preservation. Matrix spikes should not be added prior to TCLP extraction of the sample. 8.2.2 In most cases, matrix spikes should be added at a concentration equivalent to the corresponding regulatory level. If the analyte concentration is less than one half the regulatory level, the spike concentration may be as low as one half of the analyte concentration, but may not be not less than five times the method detection limit. In order to avoid differences in matrix effects, the matrix spikes must be added to the same nominal volume of TCLP extract as that which was analyzed for the unspiked sample. 8.2.3 The purpose of the matrix spike is to monitor the performance of the analytical methods used, and to determine whether matrix interferences exist. Use of other internal calibration methods, modification of the analytical methods, or use of alternate analytical methods may be needed to accurately measure the analyte concentration in the TCLP extract when the recovery of the matrix spike is below the expected analytical method performance. 8.2.4 Matrix spike recoveries are calculated by the following formula: %R (%Recovery) = 100 (X - X )/K s u where: X s = measured value for the spiked sample, X u = measured value for the unspiked sample, and K = known value of the spike in the sample. 8.3 All quality control measures described in the appropriate analytical methods shall be followed. 8.4 The use of internal calibration quantitation methods shall be employed for a metallic contaminant if: (1) Recovery of the contaminant from the TCLP extract is not at least 50% and the concentration does not exceed the regulatory level, and (2) The concentration of the contaminant measured in the extract is within 20% of the appropriate regulatory level. 8.4.1. The method of standard additions shall be employed as the internal calibration quantitation method for each metallic contaminant. 8.4.2 The method of standard additions requires preparing calibration standards in the sample matrix rather than reagent water or blank solution. It requires taking four identical aliquots of the solution and adding known amounts of standard to three of these aliquots. The forth aliquot is the unknown. Preferably, the first addition should be prepared so that the resulting concentration is approximately 50% of the expected concentration of the sample. The second and third additions should be prepared so that the concentrations are approximately 100% and CD-ROM 1311- 20 Revision 0 July 1992
150% of the expected concentration of the sample. All four aliquots are maintained at the same final volume by adding reagent water or a blank solution, and may need dilution adjustment to maintain the signals in the linear range of the instrument technique. All four aliquots are analyzed. 8.4.3 Prepare a plot, or subject data to linear regression, of instrument signals or external-calibration-derived concentrations as the dependant variable (y-axis) versus concentrations of the additions of standard as the independent variable (x-axis). Solve for the intercept of the abscissa (the independent variable, x-axis) which is the concentration in the unknown. 8.4.4 Alternately, subtract the instrumental signal or externalcalibration-derived concentration of the unknown (unspiked) sample from the instrumental signals or external-calibration-derived concentrations of the standard additions. Plot or subject to linear regression of the corrected instrument signals or external-calibration-derived concentrations as the dependant variable versus the independent variable. Derive concentrations for unknowns using the internal calibration curve as if it were an external calibration curve. 8.5 Samples must undergo TCLP extraction within the following time periods: SAMPLE MAXIMUM HOLDING TIMES [DAYS] From: From: From: Field TCLP Preparative collection extraction extraction To: To: To: Total TCLP Preparative Determinative elapsed extraction extraction analysis time Volatiles 14 NA 14 28 Semi-volatiles 14 7 40 61 Mercury 28 NA 28 56 Metals, except 180 NA 180 360 mercury NA = Not applicable If sample holding times are exceeded, the values obtained will be considered minimal concentrations. Exceeding the holding time is not acceptable in establishing that a waste does not exceed the regulatory level. Exceeding the holding time will not invalidate characterization if the waste exceeds the regulatory level. CD-ROM 1311- 21 Revision 0 July 1992
- Page 297 and 298: Table 1: Analytical Methods for the
- Page 299 and 300: All wastes received will have been
- Page 301 and 302: Waste will not be accepted until th
- Page 303 and 304: Table 3: Waste Sampling Matrix Type
- Page 305 and 306: 9 LDR TREATMENT STANDARDS Waste Des
- Page 307 and 308: Drummed consolidation waste will be
- Page 309 and 310: eevaluated whenever the process tha
- Page 311 and 312: ATTACHMENT 2-A-1 Profile Instructio
- Page 313 and 314: PMR LLC 460 W. 50 North Suite 500 S
- Page 315 and 316: • Is the waste pyrophoric? a See
- Page 317 and 318: EPA's Chemical Compatibility Chart
- Page 319 and 320: 1.0 SCOPE AND APPLICATION METHOD 90
- Page 321 and 322: 7.2.3 Adjust the electrodes in the
- Page 323 and 324: METHOD 9045D SOIL AND WASTE pH 9045
- Page 325 and 326: NOTE: Combination electrodes incorp
- Page 327 and 328: 10.0 REFERENCES 1. National Bureau
- Page 329 and 330: 1.0 SCOPE AND APPLICATION METHOD 13
- Page 331 and 332: 4.2.2 Bottle Extraction Vessel. Whe
- Page 333 and 334: 4.10 Watchglass, appropriate diamet
- Page 335 and 336: 7.0 PROCEDURE 7.1 Preliminary Evalu
- Page 337 and 338: determine whether the solid materia
- Page 339 and 340: solids content of the waste sample
- Page 341 and 342: material into an extractor bottle.
- Page 343 and 344: sample from which no additional liq
- Page 345 and 346: slowly out of the ZHE device into a
- Page 347: ® TEDLAR bag is used, if the extra
- Page 351 and 352: None of the parameters had a signif
- Page 353 and 354: Table 2. Suitable Rotary Agitation
- Page 355 and 356: Table 4. Suitable Filter Holders 1
- Page 357 and 358: Table 6. Multi-Laboratory TCLP Meta
- Page 359 and 360: Table 8. Multi-Laboratory Semi-Vola
- Page 361 and 362: Figure 1. Rotary Agitation Apparatu
- Page 363 and 364: METHOD 1311 (CONTINUED) TOXICITY CH
- Page 365 and 366: Designation: D4979 − 08 Standard
- Page 367 and 368: Designation: D4978 − 95 (Reapprov
- Page 369 and 370: FIG. 1 Method B, Sulfide Test Appar
- Page 371 and 372: TABLE 1 Determination of Bias, Pycn
- Page 373 and 374: for four known specific gravity lev
- Page 375 and 376: 21.7.1 To check the precision of sa
- Page 377 and 378: Designation: D4981 − 08 Standard
- Page 379 and 380: Designation: D5058 − 12 Standard
- Page 381 and 382: 11.6.1 Compare the temperature here
- Page 383 and 384: APPENDIX 2-E Characterization Data
- Page 385 and 386: 5. Fire-fighting measures General I
- Page 387 and 388: Carcinogenicity: CAS# 10112‐91‐
- Page 393 and 394: Combined Hazen Solution Analysis pH
- Page 395 and 396: irritation and possible burns. May
- Page 397 and 398: OSHA Vacated PELs: Mercury: 0.05 mg
150% of the expected concentration of the sample. All four aliquots are<br />
maintained at the same final volume by adding reagent water or a blank<br />
solution, and may need dilution adjustment to maintain the signals in the<br />
linear range of the instrument technique. All four aliquots are analyzed.<br />
8.4.3 Prepare a plot, or subject data to linear regression, of<br />
instrument signals or external-calibration-derived concentrations as the<br />
dependant variable (y-axis) versus concentrations of the additions of<br />
standard as the independent variable (x-axis). Solve for the intercept of<br />
the abscissa (the independent variable, x-axis) which is the concentration<br />
in the unknown.<br />
8.4.4 Alternately, subtract the instrumental signal or externalcalibration-derived<br />
concentration of the unknown (unspiked) sample from<br />
the instrumental signals or external-calibration-derived concentrations of<br />
the standard additions. Plot or subject to linear regression of the<br />
corrected instrument signals or external-calibration-derived concentrations<br />
as the dependant variable versus the independent variable. Derive<br />
concentrations for unknowns using the internal calibration curve as if it<br />
were an external calibration curve.<br />
8.5 Samples must undergo TCLP extraction within the following time<br />
periods:<br />
SAMPLE MAXIMUM HOLDING TIMES [DAYS]<br />
From: From: From:<br />
Field TCLP Preparative<br />
collection extraction extraction<br />
To: To: To: Total<br />
TCLP Preparative Determinative elapsed<br />
extraction extraction analysis time<br />
Volatiles 14 NA 14 28<br />
Semi-volatiles 14 7 40 61<br />
Mercury 28 NA 28 56<br />
<strong>Metals</strong>, except 180 NA 180 360<br />
mercury<br />
NA = Not applicable<br />
If sample holding times are exceeded, the values obtained will be considered<br />
minimal concentrations. Exceeding the holding time is not acceptable in<br />
establishing that a waste does not exceed the regulatory level. Exceeding the<br />
holding time will not invalidate characterization if the waste exceeds the<br />
regulatory level.<br />
CD-ROM 1311- 21 Revision 0<br />
July 1992