Processing kodak motion picture films, module 3 analytical procedures
Processing kodak motion picture films, module 3 analytical procedures
Processing kodak motion picture films, module 3 analytical procedures
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Quartz Cell Check<br />
Perform the quartz cell check whenever a<br />
spectrophotometric method is used that does not require the<br />
measurement of a blank.<br />
1. Inspect the cell for cracks, chips, and discoloration. If<br />
any are found, use another cell.<br />
2. Rinse the cell with reagent water several times.<br />
3. Fill the cell with reagent water.<br />
a. Wipe all outside surfaces of the cell completely<br />
dry with optical lens tissue, making sure all<br />
traces of lint are removed.<br />
b. Do not touch either of the optical faces with bare<br />
fingers.<br />
c. Look through the cell toward a light. Make sure<br />
nothing is in the water or on the optical surfaces<br />
of the cell (this includes tiny air bubbles, streaks,<br />
water spots, lint, fingerprints and anything else<br />
that might create an imperfection in the optical<br />
path).<br />
4. Set the wavelength of the spectrophotometer to<br />
240 nm for the ultraviolet range, or 650 nm for the<br />
visible range.<br />
a. Make sure that the cell holder is positioned<br />
correctly in the sample compartment.<br />
5. Adjust the spectrophotometer to read an absorbance of<br />
0 at the set wavelength.<br />
6. Insert the cell in the cell holder.<br />
a. The correct cell position is determined in the<br />
initial cell check. There is usually a difference in<br />
absorbance if the cell is placed in the holder in a<br />
reversed position.<br />
Rotating the cell 180° in its holder should not<br />
give an absorbance difference greater than<br />
0.005.1<br />
b. The cell must fit firmly and squarely into the<br />
holder (no weak or misaligned springs or<br />
corrosion).<br />
c. The optical surfaces of the cell must be perfectly<br />
aligned (at right angles) with the light beam of<br />
the spectrophotometer.<br />
7. Measure and record the absorbance reading.<br />
8. The absorbance for each wavelength should be<br />
≤ 0.093.<br />
9. If the control limits exceed 0.093, do the following to<br />
clean the cell:<br />
a. Soak the cell in water or a mild sulfonic<br />
detergent.<br />
b. If residue persists, soak the cell in a mixture of:<br />
one volume of concentrate hydrochloric acid,<br />
three volumes of reagent water, and four<br />
volumes of methanol.<br />
Caution<br />
Prepare and use this mixture only in an adequate fume hood.<br />
c. Alkaline solutions, detergents containing optical<br />
bleaches, abrasive powders, fluorides, and<br />
materials that might etch optical windows<br />
should be avoided.<br />
10. Rinse the cell thoroughly with reagent water.<br />
11. If the reading is still high, check the purity of the<br />
reagent water or use another quartz cell.<br />
Note:<br />
Do not store a cell in reagent water; let air dry.<br />
Do not use swabs or other mechanical devices to<br />
remove surface contamination; use acid/alcohol<br />
cleaning solution.<br />
Sample Measurement<br />
1. Perform the quartz cell check procedure, if the method<br />
does not require a blank. If it does involve the<br />
measurement of a blank, make sure that the cell is<br />
clean.<br />
2. Follow steps 1 - 7 of the Quartz Cell Check, but<br />
substitute the sample for reagent water and use the<br />
wavelength indicated in the specific method (step 4 of<br />
the Quartz Cell Check).<br />
3. Rinse the cell with reagent water.<br />
4. Measure the absorbance of the blank, if the method<br />
requires one.<br />
5. Using the appropriate calculation or table, determine<br />
the concentration of the component being measured.<br />
REFERENCES<br />
1. Standard Practice for Describing and Measuring<br />
Performance of Ultraviolet, Visible, and Near-<br />
Infrared Spectrophotometers; American Society for<br />
Testing and materials: ASTM designation E 275-93,<br />
Philadelphia, PA, February 1994.<br />
2. Standard Terminology Relating to molecular<br />
Spectroscopy; American Society for Testing and<br />
materials: ASTM designation E 131-94, Philadelphia,<br />
PA, August 1994.<br />
2 <strong>Processing</strong> KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03