22.07.2013 Views

Processing kodak motion picture films, module 3 analytical procedures

Processing kodak motion picture films, module 3 analytical procedures

Processing kodak motion picture films, module 3 analytical procedures

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Instructions for Performance Checks of Ultraviolet-Visible Spectrophotometers<br />

and Quartz Cells<br />

ULM-0001/1<br />

Process ECN-2 ECP-2D VNF-1/LC RVNP<br />

Formulas All All All All<br />

INTRODUCTION<br />

Performance checks for ultraviolet-visible<br />

spectrophotometers consist of a wavelength and photometric<br />

accuracy check. The visible and ultraviolet regions are<br />

checked for wavelength accuracy at 485.8 nm and 278.7 nm,<br />

respectively, with a holmium oxide solution (NIST Standard<br />

Reference Material 2034). A holmium oxide glass filter<br />

(Corning No. 3130) can be used to check wavelength<br />

accuracy at 453.2 nm and 279.4 nm, instead of the holmium<br />

oxide solution. The photometric accuracy is checked in the<br />

visible and ultraviolet region at 500 nm and 280 nm,<br />

respectively, with a metal-on-quartz filter (NIST Standard<br />

Reference Material 2031).<br />

Absorption cells should only serve as a holder for the sample<br />

and not contribute to the measured absorbance. There are<br />

almost always differences between cells. Cells should<br />

always be placed in the cell holder the same way and the<br />

holder placed the same way in the instrument. Mechanical<br />

repeatability should not contribute significant error to the<br />

<strong>analytical</strong> procedure. The most common cause for<br />

significant differences between absorption cells is dirty<br />

windows. To test cleanliness, a quartz cell check is required<br />

for spectrophotometric methods that do not involve the<br />

measurement of a blank. The quartz cell is filled with reagent<br />

water and an absorbance reading against air is taken at<br />

650 nm in the visible region and 240 nm in the ultraviolet<br />

region. The acceptable absorbance of ≤ 0.093 for each of<br />

these wavelengths are taken from ASTM E 275-931. They<br />

serve as useful control levels for analyses that use<br />

precalibrated equations in the calculations.<br />

Record all data from the performance checks on trend or<br />

control charts.<br />

APPARATUS<br />

1 cm Quartz Cell<br />

Metal-on-Quartz Filter (NIST Standard Reference<br />

Materia 2031)<br />

Holmium Oxide Solution in Ampouled Cell (NIST<br />

Standard Reference Material 2034)<br />

Holmium Oxide Glass Filter (Corning No. 3130)<br />

REAGENTS<br />

Water, Type I Reagent- This method uses reagent water<br />

equivalent to or purer than Type I grade, as defined in<br />

ASTM Standard D 1193.<br />

Acid-Alcohol Cleaning Solution- One part 3 N<br />

hydrochloric acid and one part 3A or isopropyl alcohol.<br />

PROCEDURE<br />

Preparation of Spectrophotometer<br />

1. Follow the manufacturer's instructions for operating<br />

the instrument, including the warm-up times for the<br />

instrument and it’s lamps.<br />

2. Use the appropriate lamp, as indicated:<br />

Wavelength Range Lamp<br />

200 - 380 nm Hydrogen or Deuterium<br />

380 - 780 nm Tungsten<br />

Caution<br />

The hydrogen or deuterium lamp can produce a severe burn<br />

on skin or to unprotected eyes. Do not look directly at the<br />

lamp. If it is necessary to examine or adjust the lamp, use<br />

goggles that will absorb the dangerous ultraviolet radiation.<br />

Spectrophotometer Checks<br />

Wavelength Check in Visible and Ultraviolet Region<br />

Perform this procedure at least once a month, or if the<br />

photometric accuracy check fails.<br />

1. Check the ampouled cell containing the holmium<br />

oxide solution for surface contamination; wipe clean<br />

with an optical lens tissue.<br />

2. Insert the ampouled cell into the sample holder of the<br />

spectrophotometer after zeroing the instrument from<br />

200 nm – 780 nm (air vs. air).<br />

3. Scan from 200 nm – 780 nm using the wavelength<br />

control (monochromator) in 0.5 nm increments.<br />

4. Record the wavelengths at which the maximum<br />

absorbance is observed near 485.8 nm and 278.7 nm.<br />

The recorded wavelengths should be within ± 0.5 nm<br />

of these expected values. If they are not, then contact<br />

the instrument manufacturer.<br />

Photometric Accuracy Check in Visible and<br />

Ultraviolet Region<br />

Perform this procedure at least once a day.<br />

1. Adjust the spectrophotometer to read an absorbance of<br />

0 at 280 nm and 500 nm (air vs. air).<br />

2. Check the metal on quartz filter for surface<br />

contamination; wipe clean with an optical lens tissue.<br />

3. Place the filter in sample holder.<br />

4. Read and record the absorbance value at 280 nm and<br />

500 nm.<br />

5. The absorbances at each of these wavelengths should<br />

be within trend or control chart limits established for<br />

the spectrophotometer. If they are not, then do a<br />

wavelength accuracy check as described in<br />

Wavelength Check in Visible and Ultraviolet Region.<br />

<strong>Processing</strong> KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03 1

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!