Processing kodak motion picture films, module 3 analytical procedures
Processing kodak motion picture films, module 3 analytical procedures
Processing kodak motion picture films, module 3 analytical procedures
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
4. Graduated Cylinders and Tip-up Pipets<br />
In many cases the volume of a solution to be used in an<br />
<strong>analytical</strong> method need be only an approximation of<br />
the specified volume. For example, the method may<br />
prescribe 20 mL of a reagent whereas only slightly<br />
more than 15 mL would suffice. In these cases,<br />
graduated cylinders or “tip-up” pipets are used.<br />
Graduated cylinders are calibrated to delivery (TD) or<br />
to contain (TC). See Table 1 for tolerances. “Tip-up”<br />
pipets require less time for operation and are.<br />
therefore, preferable. Portable “tip-up” pipets, shown<br />
in Figure 6, are available in different sizes. They have<br />
the accuracy of a graduated cylinder. They may be<br />
purchased with either standard taper, ground glass,<br />
male joints, or for use with rubber stoppers. It is<br />
suggested that Erlenmeyer flasks of 250-, 500-, or<br />
l000-mL capacity be selected with the ground glass<br />
joint to match. The restraining wires as shown in<br />
Figure 6 need not be used.<br />
Figure 6 Tip-up Pipet<br />
F002_0910AC<br />
5. Burets<br />
Burets are graduated to deliver variable known<br />
volumes of liquids. Methods are generally developed<br />
to employ 30 to 50 mL of solution as measured from a<br />
buret. For such purposes, a 50-mL capacity buret is<br />
used. In those cases in which 10 to 20 mL of solution<br />
are measured from a buret, a 15 or 25 mL buret is used.<br />
Class A burets are required. See Table 1 for tolerances.<br />
Analysts should understand and apply the following<br />
specific instructions for the use of burets.<br />
a. Cleanliness<br />
Use a clean buret. The buret does not have to be<br />
dry before rinsing it with the solution to be used.<br />
If the buret is not perfectly clean, drops of the<br />
solution will adhere to the wall and the buret will<br />
deliver less than the indicated volume.<br />
Furthermore, the contaminants may affect the<br />
results.<br />
b. Perfect Tip<br />
Use a buret with a good tip. A buret with a<br />
broken tip may deliver a volume other than the<br />
rated volume when the tip is touched against the<br />
wall of the receiving vessel. A buret with a<br />
broken or chipped tip sometimes can be firepolished<br />
and salvaged.<br />
c. Stopcock Seal<br />
Use a buret which will hold a constant reading<br />
for at least 5 minutes. If the stopcock seal is<br />
defective, the solution will leak and thus lower<br />
the buret reading. Teflon stopcocks do not<br />
require grease, and are preferred.<br />
d. Rinsing<br />
Rinse the entire inner surface of the buret two or<br />
three times with portions of the solution to be<br />
used.<br />
e. Lowering Meniscus to the Zero Mark<br />
Fill the buret well above the zero mark. With the<br />
buret zero mark at eye level, lower the meniscus<br />
to the zero mark. Allow a minute or two for<br />
drainage, then make the initial reading, or<br />
readjust the buret precisely to the zero mark.<br />
During the waiting period, check for leaks and<br />
make certain that air bubbles are expelled either<br />
at the top or from the tip. After the meniscus has<br />
been adjusted, remove the final drop by touching<br />
the tip with the wall of a waste-solution beaker<br />
which is kept under the buret except during<br />
titration. One-second contact is adequate.<br />
f. Position<br />
The buret should be clamped in a vertical<br />
position during the readings and while the<br />
solution is being titrated.<br />
<strong>Processing</strong> KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03 5