Processing kodak motion picture films, module 3 analytical procedures
Processing kodak motion picture films, module 3 analytical procedures
Processing kodak motion picture films, module 3 analytical procedures
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
APPARATUS<br />
Spectrophotometer with a tungsten lamp<br />
1-cm Silica Cell<br />
All volumetric glassware should meet all Class A<br />
specifications, as defined by American Society for Testing<br />
and Materials (ASTM) Standards E 287, E 288, and E 969,<br />
unless otherwise stated.<br />
REAGENTS<br />
All reagents should be ACS Reagent Grade unless otherwise<br />
specified.<br />
KODAK Anti-Calcium, No. 4<br />
Molybdenum Reagent<br />
1.0 N Sulfuric Acid, H 2 SO 4<br />
PROCEDURE<br />
A. Treatment of Sample<br />
1. Add, from a tip-up Pipet, 5 mL of KODAK Anti-<br />
Calcium No. 4 to a 100 mL volumetric flask<br />
containing a magnetic stir bar. (Allow 15 seconds for<br />
the tip-up pipet to drain.)<br />
2. Start stirring moderately. Pipet (wipe the pipet before<br />
leveling) 3.00 mL of Color Developer solution into<br />
the flask.<br />
3. Set, but do not start, a timer for five minutes.<br />
4. Add, from a tip-up pipet, 10 mL of molybdenum<br />
reagent to the flask. Immediately start the timer.<br />
5. To the flask, add, from a graduated cylinder, 80 mL of<br />
1.0 N sulfuric acid that is at room temperature<br />
(20.0°C-24.0°C).<br />
6. Remove the stirring bar and dilute the solution to its<br />
final 100.0 mL volume with 1.0 N sulfuric acid.<br />
7. Stopper the flask and invert it five times to complete<br />
mixing.<br />
B Absorbance Measurement<br />
1. Zero the spectrophotometer vs. air.<br />
2. After four minutes have elapsed, rinse and fill a 1-cm<br />
silica cell with the molybdenum blue solution<br />
(Step A.7, above).<br />
3. Place the cell in the sample compartment of the<br />
spectrophotometer and measure the solution<br />
absorbance at 725-nm (AR) at exactly five minutes<br />
after the addition of the molybdenum reagent in<br />
Step A.4, above. Record the absorbance.<br />
Note: Do not place the cell in the spectrophotometer for<br />
longer than one minute.<br />
CALCULATIONS<br />
y=mx+b<br />
Definition of the equation is found in the Regression section<br />
of Appendix A.<br />
Where:<br />
mg/L RA-1 = m (A 725 )+b<br />
Each laboratory should establish its own regression equation<br />
based on a set of calibration standards. Appendix A explains<br />
this calibration procedure. The regression equation may be<br />
different for each spectrophotometer. A typical regression<br />
equation line for the effluent is described by the following<br />
equation:<br />
Example:<br />
m = slope of the calibration line<br />
[(mg/L) RA-1 / ABU @725 ]<br />
A 725 = Absorbance of sample at 725 nm<br />
b = the intercept of the calibration line<br />
with the y-axis (mg/L RA-1)<br />
mg/L RA-1 = 143.3 (A 725 ) – 4.355<br />
A 725 = 0.847<br />
mg/L RA-1 = 143.3 (0.847) - 4.355<br />
= 117.0<br />
<strong>Processing</strong> KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03 3