Processing kodak motion picture films, module 3 analytical procedures

Processing kodak motion picture films, module 3 analytical procedures Processing kodak motion picture films, module 3 analytical procedures

motion.kodak.com
from motion.kodak.com More from this publisher
22.07.2013 Views

APPARATUS All volumetric glassware should meet all “Class A” specifications, as defined by the American Society for Testing and Materials (ASTM) Standards E 287, E 288, and E 969 unless otherwise stated. Manual Titration pH Meter, ORION EA 940, or equivalent pH Indicator electrode, CORNING Model 476024, or equivalent Calomel reference electrode filled with 3.5 N KCl, CORNING Model 476002, or equivalent 50-mL Buret Automatic Titration Titrator, METROHM 682, or equivalent pH Indicator electrode, CORNING Model 476024, or equivalent Calomel reference electrode filled with 3.5 N KCl, CORNING Model 476002, or equivalent REAGENTS All reagents used are ACS Reagent Grade unless otherwise stated. Water, Type I Reagent – This method was developed using reagent water equivalent to purer than Type I Grade, as defined in ASTM Standard D 1193. Other grades of water, e.g., reverse osmosis (RO), demineralized, or distilled water, may give equivalent results, but the effects of water quality on method performance have not been studied. 0.1 N Sulfuric Acid, standardized to four decimal places. If an autotitration system is being used, then the sulfuric acid should be standardized by using an automated system. PROCEDURE Manual Titration Titration of the Sample 1. Using Method ULM-191-2, pH Measurement of Photographic Processing Solutions (or any subsequent pH method), calibrate the pH meter for making pH measurements below pH 7. 2. Pipet (wipe the pipet before leveling) 5.0 mL of sample into a 150-mL beaker containing 50 mL of reagent water and a Teflon-coated stir bar. 3. Rinse the electrodes with reagent water and blot the electrodes with a tissue. 4. Place the electrode assembly and the tip of the buret (if possible) into the solution. Turn on the stirrer. 5. Using a 50-mL buret, titrate the sample solution with 0.1000 N sulfuric acid to a pH of 4.3. a. Add 1 mL increments of sulfuric acid to the sample solution until the pH of the solution is 5. b. Add 0.1 mL of sulfuric acid to the sample solution. When the pH of the solution reaches 4.5, record the volume of acid corresponding to the pH value. Continue to titrate until the pH of solution is less than 4.3 and record the pH after each 0.1 mL addition. c. For the calculation of total alkalinity, use the volume of acid that resulted in the solution pH that was closest to 4.300. For example, if 29.9 mL of acid resulted in a pH of 4.309, and 30.0 mL of acid resulted in a pH of 4.297, use 30.0 mL of acid in the calculation. Calculations 1. To calculate the total alkalinity (TA), use the following formula: Where: TA = (N acid )(mL acid )/0.1 Nacid = actual normality of sulfuric acid used mLacid = millilitres of sulfuric acid used 0.1 = corrects the actual normality of sulfuric acid to 0.1000 N sulfuric acid 2. The total alkalinity is reported to a tenth of a millilitre (0.1 mL). Note: The sample size used to determine the total alkalinity of a solution must be included when reporting results. For example, the total alkalinity of solution X is 24.5 mL for a 5.0 mL sample size. 2 Processing KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03

Automated Titration An example of a program listing for the determination of total alkalinity using a METROHM 682 titrator is shown in APPENDIX 1. pH Calibration Procedure 1. Fill a beaker half-full with pH 7 calibrating buffer and a second beaker half-full with pH 4 calibrating buffer. 2. Place the electrodes into a constantly stirred pH 7 calibrating buffer. 3. Press the “MEAS” key on the titrator panel. The display will read, “MEAS pH***.” 4. Press the “PREP STEP” key until the display reads “EL CAL 0/1.” 5. Enter “1” and press the “ENTER” key until the display reads “pH(s)1 7.00.” 6. Allow the electrodes to equilibrate for 2 minutes, then enter the assigned value of the pH 7 calibrating buffer and press the “ENTER” key. 7. Remove the electrodes from the cup, rinse the electrodes with reagent water, blot with a tissue, and place the electrodes into the pH 4 calibrating buffer. 8. Type in 4.00. 9. Press the “ENTER” key until the display reads “pH(s)2 4.00.” 10. Allow the electrodes to equilibrate. The slope will be printed out. The slope should be between 98 and 102 percent. If not, repeat steps 2 through 10. 11. Press the “QUIT” key to continue with the procedure. Titration of the Sample 1. Pipet (wipe the pipet before leveling) 5.0 mL of sample into a 150-mL beaker containing 50 mL of reagent water and a Teflon-coated stir bar. 2. Rinse the electrodes with reagent water and blot the electrodes with a tissue. 3. Place the electrode assembly and the tip of the buret (if possible) into the solution. turn on the stirrer. 4. Press “GO.” Calculations 1. The total alkalinity results are printed on the titrator’s printer. 2. The total alkalinity is reported to a tenth of a millilitre (0.1 mL). Note: The sample size used to determine the total alkalinity of a solution must be included when reporting results. For example, the total alkalinity of solution X is 24.5 mL for a 5.0 mL sample size. APPENDIX 1 METROHM 682 Titrator Example Programs Total Alkalinity Determination Set pH 4.3 Prep. Steps Pause 10 s Titr.dosimat 1 Electr. input 1 Parameters: EP1 pH 4.30 Dyn.pH 1 3.0 Drift1 10.0 mV/s T(delay) 1 s Temp. 25°C Stop V 99.99 mL Formula: F1 = (F1 = (EP1*C01) Formula Constant: C01 = correction factor for the normality of the sulfuric acid. C01 can be calculated as follows: Actual normality of sulfuric acid 0.1000 Processing KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03 3

APPARATUS<br />

All volumetric glassware should meet all “Class A”<br />

specifications, as defined by the American Society for<br />

Testing and Materials (ASTM) Standards E 287, E 288, and<br />

E 969 unless otherwise stated.<br />

Manual Titration<br />

pH Meter, ORION EA 940, or equivalent<br />

pH Indicator electrode, CORNING Model 476024, or<br />

equivalent<br />

Calomel reference electrode filled with 3.5 N KCl,<br />

CORNING Model 476002, or equivalent<br />

50-mL Buret<br />

Automatic Titration<br />

Titrator, METROHM 682, or equivalent<br />

pH Indicator electrode, CORNING Model 476024, or<br />

equivalent<br />

Calomel reference electrode filled with 3.5 N KCl,<br />

CORNING Model 476002, or equivalent<br />

REAGENTS<br />

All reagents used are ACS Reagent Grade unless otherwise<br />

stated.<br />

Water, Type I Reagent – This method was developed<br />

using reagent water equivalent to purer than Type I Grade,<br />

as defined in ASTM Standard D 1193. Other grades of<br />

water, e.g., reverse osmosis (RO), demineralized, or<br />

distilled water, may give equivalent results, but the effects<br />

of water quality on method performance have not been<br />

studied.<br />

0.1 N Sulfuric Acid, standardized to four decimal places.<br />

If an autotitration system is being used, then the sulfuric<br />

acid should be standardized by using an automated<br />

system.<br />

PROCEDURE<br />

Manual Titration<br />

Titration of the Sample<br />

1. Using Method ULM-191-2, pH Measurement of<br />

Photographic <strong>Processing</strong> Solutions (or any<br />

subsequent pH method), calibrate the pH meter for<br />

making pH measurements below pH 7.<br />

2. Pipet (wipe the pipet before leveling) 5.0 mL of<br />

sample into a 150-mL beaker containing 50 mL of<br />

reagent water and a Teflon-coated stir bar.<br />

3. Rinse the electrodes with reagent water and blot the<br />

electrodes with a tissue.<br />

4. Place the electrode assembly and the tip of the buret (if<br />

possible) into the solution. Turn on the stirrer.<br />

5. Using a 50-mL buret, titrate the sample solution with<br />

0.1000 N sulfuric acid to a pH of 4.3.<br />

a. Add 1 mL increments of sulfuric acid to the<br />

sample solution until the pH of the solution is 5.<br />

b. Add 0.1 mL of sulfuric acid to the sample<br />

solution. When the pH of the solution reaches<br />

4.5, record the volume of acid corresponding to<br />

the pH value. Continue to titrate until the pH of<br />

solution is less than 4.3 and record the pH after<br />

each 0.1 mL addition.<br />

c. For the calculation of total alkalinity, use the<br />

volume of acid that resulted in the solution pH<br />

that was closest to 4.300. For example, if<br />

29.9 mL of acid resulted in a pH of 4.309, and<br />

30.0 mL of acid resulted in a pH of 4.297, use<br />

30.0 mL of acid in the calculation.<br />

Calculations<br />

1. To calculate the total alkalinity (TA), use the<br />

following formula:<br />

Where:<br />

TA = (N acid )(mL acid )/0.1<br />

Nacid = actual normality of sulfuric acid used<br />

mLacid = millilitres of sulfuric acid used<br />

0.1 = corrects the actual normality of sulfuric acid<br />

to 0.1000 N sulfuric acid<br />

2. The total alkalinity is reported to a tenth of a millilitre<br />

(0.1 mL).<br />

Note: The sample size used to determine the total<br />

alkalinity of a solution must be included when<br />

reporting results. For example, the total alkalinity of<br />

solution X is 24.5 mL for a 5.0 mL sample size.<br />

2 <strong>Processing</strong> KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!