Potentiometric Determination of Silver in Process ECN Fixing Baths<br />

ECN-0023-01<br />

INTRODUCTION<br />

The sample containing silver is titrated potentiometrically<br />

with standardized sodium sulfide using a silver billet/double<br />

junction electrode pair. The sample is made alkaline to<br />

prevent the decomposition of sodium thiosulfate, which<br />

occurs in acid solutions. Ethylenedinitrilotetraacetic acid<br />

(EDTA) is added to minimize interference of other metal<br />

ions. The EDTA reagent does not prevent interference from<br />

zinc ions. Gelatin is added to prevent the coagulation of the<br />

silver sulfide that is formed. This prevents the coagulated<br />

silver sulfide from occluding the silver ions.<br />

Changes in the volume of sample and of sodium<br />

hydroxide/EDTA reagent affect the silver results. If a small<br />

amount of sample is used, the sample volume must be<br />

adjusted to about 300 mL with 1.0 M sodium thiosulfate.<br />

Use of this method requires handling potentially<br />

hazardous chemicals. Material Safety Data Sheets (MSDS)<br />

should be consulted for each chemical before use. These can<br />

be obtained from each chemical supplier.<br />

PRECISION AND BIAS<br />

A. Repeatability Standard Deviation, 1sr and<br />

95 Percent Confidence Estimate (not including<br />

Bias)<br />

Repeatability standard deviation is an estimate of the<br />

variability one trained analyst should be able to obtain under<br />

favorable conditions (analyzing a sample, with one<br />

instrument, within one day).<br />

The 95 percent confidence estimate (calculated using the<br />

repeatability standard deviation) around a single test result<br />

will include the mean value 95 percent of the time.<br />

To obtain the repeatability data, a single skilled analyst<br />

performed three (3) replicates on each of the following<br />

solutions during methods development.<br />

1. A “fresh” fixing tank prepared with all components at<br />

their respective “working tank” aim concentrations.<br />

2. The same “fresh” solution as in number 1, above,<br />

reanalyzed after making an <strong>analytical</strong>ly weighed,<br />

standard addition of 1.0745 g/L Ag.<br />

3. A “seasoned” fixing tank analyzed potentiometrically<br />

as received, at 1.1488 g/L Ag.<br />

4. The same “seasoned” solution as in number 3, above,<br />

reanalyzed after making an <strong>analytical</strong>ly weighed,<br />

standard addition of 0.2686 g/L Ag.<br />

Sample<br />

Mean<br />

(g/L Ag)<br />

Bias<br />

Bias is a statistically significant deviation of the mean from<br />

the known mix level at a 95 percent confidence level. It is<br />

determined for fresh samples only. Bias was not determined<br />

for this sample because the component concentration level<br />

was not determined independently of the test method.<br />

Instead, a recovery was calculated for the component in a<br />

fresh sample.<br />

The recovery of 106.23 percent was not significantly<br />

different from 100 percent at the 95 percent confidence<br />

level.<br />

Recovery<br />

Recovery is used for seasoned samples, since the component<br />

concentration level was not determined independently of the<br />

test method. It is defined as the calculated mean for the<br />

seasoned sample with a standard addition of the component<br />

minus the mean for the seasoned sample, divided by the<br />

actual amount of the standard addition. It is expressed as a<br />

percentage.<br />

The recovery of 102.16 percent* was not statistically<br />

significantly different from 100 percent at the 95 percent<br />

confidence level.<br />

* Note: Recovery was calculated by accounting for a<br />

5 percent error from dilution of 1 liter of seasoned fixer by a<br />

50.00 mL aliquot of silver nitrate.<br />

<strong>Processing</strong> KODAK Motion Picture Films, Module 3, Analytical Procedures H24.03 1<br />

Ag<br />

N<br />

Repeatability<br />

Standard<br />

Deviation, 1s r<br />

(g/L Ag)<br />

95 Percent<br />

Confidence<br />

Estimate<br />

(g/L Ag)<br />

“Fresh”<br />

“Fresh” plus<br />

0.0060 3 0.00000 ± 0.0000<br />

Standard<br />

Addition<br />

“Seasoned”,<br />

1.0830 3 0.00251 ± 0.0108<br />

As<br />

Received<br />

“Seasoned”<br />

1.1488 3 0.00374 ± 0.0161<br />

plus<br />

Standard<br />

Addition<br />

1.3658 3 0.00492 ± 0.0212<br />

Example:<br />

(1.3658) - (1.1488 x 0.95) x 100 = 102.16%<br />

0.2686