P36 - Comparison <strong>of</strong> Veratox® for T-2/HT-2 ELISA test<br />
with GC-MS and LC-MS methods<br />
S. Rämö 1 , P. Parikka 1 , A. Asola, 2 M. Jestoi 2<br />
1 MTT Agrifood Research Finland, 31600 Jokioinen, Finland and 2 Finnish Food Safety authority Evira,<br />
00790 Helsinki, Finland<br />
E-mail: sari.ramo@mtt.fi<br />
Neogen Veratox® for T-2/HT-2 toxin is ELISA test which promises 100 % cross<br />
reaction for both toxins. The usefulness <strong>of</strong> the test was studied in MTT during<br />
2011-2012. The possible matrix interferences were tested for wheat, barley and<br />
oats, which were analyzed with a GC-MS reference method: Wheat did not show<br />
any matrix interference. However, the need for matrix calibration for barley and<br />
oats was observed. T-2 and HT-2 toxin free barley and oats samples were spiked<br />
with toxin levels <strong>of</strong> 25 ppb to 250 ppb. Matrix interference was clear for barley,<br />
whereas for oats the test’s calibration could be used.<br />
T-2 and HT-2 toxins are a more considerable problem in oats than other grains in<br />
Finland. 20 oats samples, which were analyzed with LC-MS in 2011, were chosen<br />
for ELISA analysis. The ELISA tests were done during four days in August 2012.<br />
The standards and samples were transferred to wells according to the instruction<br />
<strong>of</strong> the kit. The highest calibration level is only 250 ppb and most <strong>of</strong> the sample<br />
extracts needed extra dilutions for quantitative result. The absorbances were<br />
measured with 650 nm with Tecan Infinite F200 microwell reader and Magellan<br />
s<strong>of</strong>tware. The acceptable correlation (r ≥ 0,995) <strong>of</strong> the calibration curve were not<br />
achieved in the first day, but were excellent (r = 0.999) in 2 nd and 4 th day. The<br />
variation between the absorbance values <strong>of</strong> the same calibration level was low,<br />
because <strong>of</strong> close measurements days and the same lot number <strong>of</strong> used kits.<br />
ELISA results <strong>of</strong> 20 oats samples were compared with LC-MS results. The slope<br />
<strong>of</strong> the regression line was 1.226, so the ELISA gave little bit higher results than<br />
LC-MS method. ELISA test worked quite well when sum concentration <strong>of</strong> T-2 and<br />
HT-2 toxins was 25 - 250 ppb.<br />
Keywords: T-2, HT-2, ELISA, correlation<br />