TRITON Extreme Operation Guide - Platinum Audiolab
TRITON Extreme Operation Guide - Platinum Audiolab
TRITON Extreme Operation Guide - Platinum Audiolab
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
76<br />
E Notice that 0000: LOOP1-140BPM-L is selected for<br />
“Sample (Sample Select),” and that the sample waveform<br />
data is displayed.<br />
The selected sample is a stereo sample. The L and R<br />
waveforms will be shown in the upper and lower parts<br />
of the sample waveform display.<br />
F Use “Start” to set the start address, “LoopS (Loop<br />
Start)” to set the loop start address, and “End” to set<br />
the end address.<br />
The sample will sound as follows.<br />
When looping is on: S → E → LoopS → E → (continue<br />
repeating LoopS → E)<br />
When looping is off: S → E<br />
Select “Start” (highlighted), and use the [VALUE] dial<br />
etc. to specify the location at which the sample will<br />
begin sounding. The vertical line in the display will<br />
move accordingly.<br />
If necessary, you can press the ZOOM buttons to<br />
expand or shrink the waveform display. Zoom will<br />
expand or contract the display beginning at the<br />
“Start,” “LoopS (Loop Start),” or “End” point that is<br />
selected.<br />
Zoom-in vertically<br />
Zoom-out horizontally Zoom-in horizontally<br />
Maximum<br />
horizontal zoom-out<br />
1x horizontal zoom-in<br />
Zoom-out vertically<br />
If you check “Use Zero,” locations where the waveform<br />
is at zero will automatically be found and<br />
selected when you use the [VALUE] dial etc. to set<br />
“Start,” “LoopS (Loop Start),” or “End” points. This<br />
makes it easy to set these points to addresses that will<br />
not produce clicks or noise in the loop.<br />
For this example, set “LoopS (Loop Start)” and<br />
“Start” to the same value.<br />
If desired, you can make a dotted vertical line appear<br />
in the sample waveform display to indicate the BPM.<br />
If you use this function with waveforms that have a<br />
specific BPM, it will be easier to make accurate settings<br />
for “End.”<br />
G Select the page menu command “Grid” to access the<br />
dialog box.<br />
Check “Grid,” set “Resolution” to , and press the OK<br />
button.<br />
Set “Grid” to 140. With these settings, vertical dotted<br />
lines will be displayed at 140 BPM quarter note intervals.<br />
Set “End.”<br />
When “Loop” is On, the dotted vertical “Grid” lines<br />
will start at “LoopS (Loop Start).” If you want to set<br />
“End” at the end of one 4/4 measure, the fourth vertical<br />
line from the “LoopS (Loop Start)” line will be the<br />
end of the first measure. Place “End” at this vertical<br />
line.<br />
H Use “Truncate” to delete the unwanted data that is<br />
outside of the start (or loop start) and end addresses.<br />
Select the page menu command “Truncate” to access<br />
the dialog box.<br />
Select the Front & End radio button.<br />
In this example, we will not change the settings of the<br />
“Save to No.” and “Overwrite” check boxes, so press<br />
the OK button to execute. When you execute the operation,<br />
the truncated samples 0002: LOOP1-140B0002-L<br />
and 0003: LOOP1-140B0002-R will be automatically<br />
assigned to Index 1.<br />
Please refer to “A note on saving samples” (☞p.77)<br />
for a cautionary note regarding the “Save to No.” and<br />
“Overwrite” check boxes.