6<br />

ROTOR PREPARATION<br />

Hanger<br />

Bucket<br />

<strong>28</strong> 000-rpm<br />

64-Sector<br />

(342211)<br />

Cap<br />

O-ring<br />

812715 (<strong>SW</strong> <strong>28</strong>)<br />

815472 (<strong>SW</strong> <strong>28</strong>.1)<br />

2. Make sure that the rotor is equipped with the correct overspeed<br />

disk. If the disk is missing or damaged, replace it according to the<br />

instructions in Rotors <strong>and</strong> Tubes.<br />

3. Check the chemical compatibilities of all materials used (refer to<br />

Appendix A in Rotors <strong>and</strong> Tubes).<br />

4. Verify that the tubes <strong>and</strong> bottles being used are listed in Table 2 or<br />

Table 3.<br />

For runs at other than room temperature, refrigerate or warm the<br />

rotor beforeh<strong>and</strong> for fast equilibration.<br />

Place the rotor on the rotor st<strong>and</strong> (332400) when it is not in the<br />

centrifuge.<br />

1. Load the filled containers into the buckets (see page 8 for tube<br />

information). Complete loading by placing the correct floating<br />

spacers (if required) over the tubes.<br />

2. Ensure that bucket O-rings are lightly but evenly coated with<br />

silicone vacuum grease (335148). Do not run a bucket without an<br />

O-ring, as the bucket will leak.<br />

3. Be sure that metal threads in the bucket caps are clean <strong>and</strong> lightly<br />

but evenly lubricated with Spinkote lubricant (306812). Match<br />

bucket caps with numbered buckets <strong>and</strong> screw them down<br />

manually until tight.<br />

4. Hook the buckets to the rotor by inserting the bucket pins into<br />

the grooves on the rotor body. Swing each bucket back <strong>and</strong> forth<br />

slightly to ensure proper installation; the buckets should move<br />

freely. Six buckets must be installed, whether loaded or empty.<br />

If fewer than six tubes are being run, they must be arranged<br />

symmetrically in the rotor (see Figure 1). Opposing tubes must<br />

be filled to the same level with liquid of the same density.